FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1200626 · Received October 20, 2008

Report

Report Number
6000001-2007-01540
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
December 21, 2006
Report Date
December 21, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16, 2007. EVALUATION SUMMARY: THE DEVICE EVALUATION WAS COMPLETED AND THE DEPLETED BATTERY CONDITION WAS CONFIRMED DUE POTENTIALLY DAMAGED BATTERIES. BAXTER SERVICE TECHNICIAN INSTALLED NEW BATTERIES AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE DEVICE WAS RECEIVED FOR SERVICE. DURING SERVICE TESTING DEPLETED BATTERIES WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1