COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-01269
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 19, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTION FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 11, 2007. EVALUATION SUMMARY:THE REPORTED CONDITION OF A FAILURE CODE 814:01 WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THIS FAILURE CODE WAS CAUSED BY DEPLETED BATTERIES, THEREFORE THE PUMP'S BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THIS FAILURE CODE. THIS FAILURE CODE IS BEING INVESTIGATED UNDER CAPA. THE ISSUE OF DEPLETED BATTERIES WILL BE INVESTIGATED UNDER CAPA.
THE FACILITY REPORTED A PUMP WITH FAILURE CODE 814:01. IT IS UNKNOWN WHEN THIS OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.