FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1200487 · Received October 20, 2008

Report

Report Number
6000001-2007-01776
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
December 1, 2006
Report Date
December 22, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTION FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 17, 2007. EVALUATION SUMMARY:FAILURE CODE 570:320:844:0000 WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE CAUSE OF THE FAILURE CODE WAS DUE TO DEPLETED AND POTENTIALLY DAMAGED BATTERIES. THE BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA, WHICH WAS OPENED ON APRIL 1, 2005.

Description of Event or Problem · 1

THE FACILITY REPORTED TO CUSTOMER SERVICE A PUMP WITH FAILURE CODE 570:320:844:0000. IT IS UNKNOWN WHEN THIS FAILURE CODE OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1