FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1200430 · Received October 20, 2008

Report

Report Number
6000001-2007-01864
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
January 12, 2007
Report Date
January 12, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING SUBMITTED IN ACCORDANCE WITH INSTRUCTION FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 17, 2007. EVALUATION SUMMARY: THE CONDITION OF DEPLETED BATTERIES WAS CONFIRMED. THE BATTERIES WERE REPLACED DUE TO POTENTIAL DAMAGE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE.

Description of Event or Problem · 1

THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN NOTED DEPLETED BATTERIES. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1