AEQUALIS REVERSED II UNIVERSAL GLENOID BASEPLATE IMPACTOR
Report
- Report Number
- 3000931034-2021-00280
- Event Type
- Malfunction
- Date Received
- June 15, 2021
- Date of Event
- May 13, 2021
- Report Date
- December 23, 2021
- Manufacturer
- TORNIER S.A.S.
- Product Code
- HWA
- UDI-DI
- 03700386968969
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- 003
Narratives
CORRECTIONS: PLEASE REFER TO D9/H3 PRODUCT AVAILABLE TO STRYKER ; H4 MANUFACTURING DATE ; H6 METHOD CODE, RESULTS CODE, CONCLUSION CODE, AND HEALTH IMPACT CODE. THE REPORTED EVENT COULD NOT BE CONFIRMED, SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO OTHER EVIDENCE WERE PROVIDED. THE DEVICE INSPECTION WAS NOT POSSIBLE AS THE PRODUCT WAS NOT RETURNED FOR INVESTIGATION. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. MORE DETAILED INFORMATION ABOUT THE COMPLAINT EVENT AS WELL AS THE AFFECTED DEVICE MUST BE AVAILABLE TO DETERMINE THE ROOT CAUSE OF THE COMPLAINT EVENT. IF DEVICE IS RETURNED OR ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE REASSESSED. H3 OTHER TEXT : DEVICE DISCARDED.
TIGHTENING BOLT ON INNER ROD HAD STRIPPED THREADS. MAKING IT IMPOSSIBLE TO ASSEMBLE WITH IMPACTOR HANDLE. THEREFORE MAKING IT IMPOSSIBLE TO ATTACH THE BASEPLATE TO THE IMPACTOR. SURGEON TRIED TO ASSEMBLE THE IMPACTOR WITH VISE GRIPS WITHOUT SUCCESS. NO BACK-UP DEVICE WAS AVAILABLE. THE SURGEON HAD TO HOLD THE BASEPLATE TIGHT WITH HIS FINGERS WHILE HAMMERING THE IMPACTOR. THIS CAUSED A 20 MINUTE DELAY.
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
TIGHTENING BOLT ON INNER ROD HAD STRIPPED THREADS. MAKING IT IMPOSSIBLE TO ASSEMBLE WITH IMPACTOR HANDLE. THEREFORE MAKING IT IMPOSSIBLE TO ATTACH THE BASEPLATE TO THE IMPACTOR. SURGEON TRIED TO ASSEMBLE THE IMPACTOR WITH VISE GRIPS WITHOUT SUCCESS. NO BACK-UP DEVICE WAS AVAILABLE. THE SURGEON HAD TO HOLD THE BASEPLATE TIGHT WITH HIS FINGERS WHILE HAMMERING THE IMPACTOR. THIS CAUSED A 20 MINUTE DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 902237 | AEQUALIS REVERSED II UNIVERSAL GLENOID BASEPLATE IMPACTOR | IMPACTOR | HWA | TORNIER S.A.S. | GLENOID BASEPLATE IMPACTOR | 10B803 | 03700386968969 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |