FDA Adverse Event Other Summary report: N

BIO-TRANSFIX, 5MM X 40MM, LONG

MDR report key: 1200203 · Received October 6, 2008

Report

Report Number
1220246-2008-00197
Event Type
Other
Date Received
October 6, 2008
Date of Event
September 4, 2008
Report Date
September 8, 2008
Manufacturer
ARTHREX, INC.
Product Code
MNU
PMA / PMN Number
K011172
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION REVEALED A BOW/BEND AT THE MID-POINT OF THE DEVICE. STRESS/SCUFF MARKS WERE ON THE IMPLANT DIRECTLY WHERE THE BOW/BEND WAS PRESENT; THERE WAS NO ACTUAL CRACK CONFIRMED IN THE DEVICE. IN ADDITION, THE DISTAL TIP OF THE IMPLANT WAS DEFORMED TO AN OVAL SHAPE. THE LOT NUMBER WAS REQUESTED BUT NOT PROVIDED FROM THE REPORT OF THE INITIAL SURGERY AND COULD NOT BE DETERMINED FROM DEVICE EVALUATION. DHR COULD NOT BE REVIEWED AND MANUFACTURING DATE COULD NOT BE DETERMINED SINCE THE LOT NUMBER WAS NOT PROVIDED. THE MOST LIKELY CAUSE OF THIS TYPE OF EVENT IS GRAFT WIRE THAT IS NOT INSERTED STRAIGHT. IF, DURING INSERTION, THE GRAFT WIRE IS PASSED THROUGH THE GRAFT AND IT IS NOT STRAIGHT, THE IMPLANT WILL TEND TO CONFORM TO THE SAME PATH AS THE GRAFT PASSING WIRE. CONSTANT AND/OR DYNAMIC LOAD PLACED ON THE IMPLANT AFTER FOUR MONTHS IN VIVO COULD HAVE CAUSED THE PLASTIC DEFORMATION (BOW/BEND) IN THIS CASE. IN ADDITION, WHEN THE IMPLANT WAS SEATED, THE WIRE MOST LIKELY CAUSED THE ID (INNER DIAMETER) AT THE TIP TO DEFORM AS WELL. THIS CONDITION WAS NOT DETRIMENTAL TO THE PERFORMANCE OF THE DEVICE, BUT HELPED TO CONFIRM THE EVALUATION CONCLUSION THAT THE GRAFT WIRE WAS MOST LIKELY NOT INSERTED STRAIGHT. THE POSSIBLE ROOT CAUSE OF THIS EVENT WAS DETERMINED AS MISUSE BY USER. THIS IS THE FIRST COMPLAINT OF THIS PART NUMBER. THE POTENTIAL CAUSES OF THIS EVENT ARE BEING COMMUNICATED TO THE EVENT REPORTER.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ACL REVISION SURGERY WAS PERFORMED DUE TO THE PATIENT COMPLAINING OF PAIN WHERE A TRANSFIX PIN WAS IMPLANTED. FROM THE REVISION, THE SURGEON FOUND THE IMPLANT WAS CRACKED DOWN THE MIDDLE. HE REMOVED THE PIECE THAT WAS NOT ENGAGED, LEFT THE REMAINING PIECE AS IT WAS, AND COMPLETED THE REVISION CASE. FOLLOW-UP WITH THE REPORTER PROVIDED INFORMATION THAT THE REVISION OCCURRED 4 MONTHS POST-OP; THE INITIAL IMPLANT WAS ON (B) (6) 2008 AND THE REVISION WAS ON (B) (6) 2008. THE BROKEN PIECE OF THE SCREW WAS REMOVED IN THE REVISION CASE. THE IMPLANT WAS REPORTED TO HAVE BEEN 3/4 SEATED. THE PATIENT BONE QUALITY WAS REPORTED AS AVERAGE. NO FURTHER INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIO-TRANSFIX, 5MM X 40MM, LONG MNU ARTHREX, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 32 YR Other