NXTAG COV EXTENDED PANEL
Report
- Report Number
- 3002777243-2021-00006
- Event Type
- Malfunction
- Date Received
- June 15, 2021
- Date of Event
- February 8, 2021
- Report Date
- June 15, 2021
- Manufacturer
- LUMINEX MOLECULAR DIAGNOSITICS
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- 003
Narratives
THERE WAS NO REPORTED INJURY OR DEATH AT THE TIME OF THIS REPORT. (B)(4).
CUSTOMER IS REPORTING INCONSISTENT RESULTS BETWEEN KIT LOTS. TECHNICAL SUPPORT (TS) REVIEWED RAW DATA AND FOUND: MULTIPLE DISCREPANT CALLS, HIGH BACKGROUND, VARIABLE BETWEEN THE SAMPLES, NO NEGATIVE CONTROLS (NTC) WERE BEING RUN. THIS WAS DISCUSSED WITH MOLECULAR APPLICATIONS SPECIALIST (MAS) AND IT WAS SUGGESTED TO RUN SAMPLES IN ORDER BELOW: NTC, NEGATIVE EXTRACTION CONTROL (NEC), POSITIVE CONTROL, NTC, NEC, NTC, NEC, NTC AND ADVISED TO SWITCH OVER TO DRIVE FLUID PLUS. CUSTOMER SENT SUPPORT UTILITY FILES (SUF) FOR BOTH MAGPIX UNITS AND A PREVIOUS RUN WITH NTC AND NEC WHICH WERE CLEAN. THIS WAS ALSO RUN USING THE NEW KIT IN QUESTION. IT WAS CONCLUDED THAT KIT ISSUES CAN BE RULED OUT. TS REVIEWED THE SUF FOR BOTH UNITS AND FOUND CHAMBERS WERE CLEAN. NO MAJOR ERRORS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 902205 | NXTAG COV EXTENDED PANEL | NXTAG COV EXTENDED PANEL | QJR | LUMINEX MOLECULAR DIAGNOSITICS | IK054C-0054 |