FDA Adverse Event
Malfunction
Summary report: N
ELECSYS 2010 RACK
MDR report key: 1200094
·
Received October 15, 2008
Report
- Report Number
- 1823260-2008-07615
- Event Type
- Malfunction
- Date Received
- October 15, 2008
- Date of Event
- October 7, 2008
- Report Date
- October 15, 2008
- Manufacturer
- HITACHI HIGH TECH CORP
- Product Code
- JJE
- PMA / PMN Number
- K961481
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
USER EXPERIENCED CONTINUED LEAK FROM THE ANALYZER THAT STARTED IN 2008. THE LEAK WAS REACHING ONTO THE FLOOR AND INTO THE WALKWAY. THE USER WAS NOT HARMED AND NO PATIENT SAMPLES WERE AFFECTED. THE FIELD SERVICE REPRESENTATIVE FOUND THAT A RUBBER GASKET WAS NOT SEALING CORRECTLY ON THE WATER RESERVOIR. THE GASKET WAS CLEANED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ELECSYS 2010 RACK | IMMUNOCHEMISTRY ANALYZER - JJE | JJE | HITACHI HIGH TECH CORP | 2010 RACK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |