FDA Adverse Event Injury Summary report: N

MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 32 SIZE S -4

MDR report key: 11999155 · Received June 15, 2021

Report

Report Number
3005180920-2021-00497
Event Type
Injury
Date Received
June 15, 2021
Date of Event
May 19, 2021
Report Date
June 15, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809156
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 31 MAY 2021: LOT 180530: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20-JUNE-2018. EXPIRATION DATE: 2023-06-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL DEVICES INVOLVED BATCH REVIEWS PERFORMED ON 31 MAY 2021: STEM: AMISTEM H 01.18.130 HA COATED STD STEM SIZE 0 (K093944) LOT 177141: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-APR-2018. EXPIRATION DATE: 2023-03-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. LINER: MPACT 01.32.3241HCT FLAT PE HC LINER Ø32/D (K103721) LOT 180650: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-MAY-2018. EXPIRATION DATE: 2023-04-26. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING PAIN 2 YEARS 6 MONTHS AFTER THE PRIMARY AND THE CAUSE OF PAIN IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH A COMPETITOR'S PRODUCT AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
901896 MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 32 SIZE S -4 HIP CERAMIC HEAD LZO MEDACTA INTERNATIONAL SA 01.29.204 180530 07630030809156

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention