FDA Adverse Event Malfunction Summary report: N

PICCOLO COMPOSITE PF NAILING SYSTEM - LAG SCREW

MDR report key: 11997309 · Received June 15, 2021

Report

Report Number
9615128-2021-00003
Event Type
Malfunction
Date Received
June 15, 2021
Date of Event
May 27, 2021
Report Date
June 15, 2021
Manufacturer
CARBOFIX ORTHOPEDICS LTD.
Product Code
HSB
UDI-DI
07290109337852
PMA / PMN Number
K153536
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE PRODUCTION RECORDS OF THE INVOLVED DEVICE DID NOT REVEAL A DEVIATION THAT MAY HAVE CONTRIBUTED TO THE REPORTED CASE. THE INVOLVED IMPLANT WAS NOT AVAILABLE TO THE COMPANY; THUS, ITS PHYSICAL EXAMINATION WAS NOT POSSIBLE. AN X-RAY OF THE CASE WAS PROVIDED TO THE MANUFACTURER. IT SHOULD BE NOTED, THAT THE TITANIUM SHELL IS ATTACHED TO THE LAG SCREW DURING THE MOLDING PROCESS AND THUS IN ORDER TO DISCONNECT THE TWO, HIGH TORQUE FORCES ARE REQUIRED. IN THE INVOLVED CASE, IT IS POSSIBLE THAT THE LAG SCREW WAS INSERTED TOO DEEP AND REACHED THE BONE CORTEX. THUS WHEN IT WAS UNSCREWED FOR EXTRACTION, HIGH FORCES WERE APPLIED ON THE SCREW'S THREAD LEADING TO THE SEPARATION OF THE TITANIUM SHELL.

Description of Event or Problem · 1

DURING IMPLANTATION OF A PICCOLO COMPOSITE PF NAIL IN (B)(6), AFTER LAG SCREW INSERTION AND X-RAY EXAMINATION, THE SURGEON DECIDED TO REMOVE THE LAG SCREW AND REPLACE IT WITH A SHORTER LAG SCREW SINCE IT WAS TOO LONG. AFTER LAG SCREW REMOVAL, THE SURGEON DISCOVERED THAT THE TITANIUM SHELL AT THE LAG SCREW DISTAL END SEPARATED FROM THE LAG SCREW AND REMAINED IN THE BONE. A SHORTER LAG SCREW WAS INSERTED WITH SHELL REMAINING IN SITU.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
894860 PICCOLO COMPOSITE PF NAILING SYSTEM - LAG SCREW INTRAMEDULLARY FIXATION ROD SYSTEM HSB CARBOFIX ORTHOPEDICS LTD. Lag Screw PPF01726 07290109337852

Patients

Seq Age Sex Outcome Treatment
1