VISERA ELITE II VIDEO SYSTEM CENTER
Report
- Report Number
- 8010047-2021-07508
- Event Type
- Malfunction
- Date Received
- June 14, 2021
- Report Date
- July 7, 2021
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- OWN
- PMA / PMN Number
- K200542
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE DEVICE WAS RETURNED TO OLYMPUS REPAIR CENTER, BUT NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC). THEREFORE, OMSC COULD NOT CONFIRM THE DEVICE. DEVICE HISTORY RECORD REVIEW INDICATES THAT THE PRODUCT WAS MANUFACTURED AND TESTED IN ACCORDANCE WITH ALL APPLICABLE PROCEDURES AND MET ALL FINAL PRODUCT RELEASE CRITERIA. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. OMSC SURMISED THAT THIS PHENOMENON ATTRIBUTED TO THE BROKEN FAN OF THE DEVICE BY THE USER HANDLING. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
THE DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC), BUT RETURNED TO OLYMPUS REPAIR CENTER FOR EVALUATION. ACCORDING TO THE EVALUATION, THE FOLLOWING WAS FOUND. POWER SUPPLY UNIT COOLING FANS WERE NOT ROTATING DUE TO THE FAN WAS CRACKED. REAR PANEL FAN MADE A NOISE DUE TO A FAILURE FAN UNIT. TOP COVER AND REAR PANEL WERE DENTED. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
THE COOLING FAN MOTOR OF THE DEVICE DID NOT WORK. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT. THE USER FACILITY DID NOT PROVIDE OTHER DETAILED INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 892775 | VISERA ELITE II VIDEO SYSTEM CENTER | VIDEO SYSTEM CENTER | OWN | OLYMPUS MEDICAL SYSTEMS CORP. | OTV-S200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |