FDA Adverse Event Malfunction Summary report: N

CNS-6801A

MDR report key: 11996789 · Received June 14, 2021

Report

Report Number
8030229-2021-00338
Event Type
Malfunction
Date Received
June 14, 2021
Date of Event
May 15, 2021
Report Date
February 21, 2023
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
UDI-DI
04931921131640
PMA / PMN Number
K102376
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DETAILS OF COMPLAINT: THE NURSE REPORTED THAT THE TELEMETRY TRANSMITTERS WERE IN COMM LOSS AT THE CENTRAL NURSE'S STATION (CNS). NO PATIENT HARM OR INJURY WAS REPORTED. INVESTIGATION SUMMARY: THE ROOT CAUSE FOR COMMUNICATION LOSS IS RELATED TO THIRD-PARTY DEVICES. NKA WAS ABLE TO RESOLVE THE ISSUE OF COMMUNICATION LOSS BY REPLACING DEFECTIVE NETWORK SWITCHES WHICH WERE SUPPLIED TO THE CUSTOMER BY A THIRD-PARTY. THERE IS NO EVIDENCE OF AN NK DEVICE MALFUNCTION. NO FURTHER ISSUES RELATING TO COMMUNICATION LOSS WITH THIS CNS HAVE BEEN REPORTED.

Additional Manufacturer Narrative · 0

THE NURSE REPORTED THAT THEY ARE EXPERIENCING A COMMUNICATION LOSS WITH A CENTRAL NURSE'S STATION (CNS) IN EAST AND 5 WEST. THEY STATED THERE IS CONSTRUCTION GOING ON WHICH KNOCKED OUT POWER TO NK DEVICES AND OTHER (B)(6) OWNED EQUIPMENT EARLIER IN THE DAY. POWER WAS RESTORED, BUT CNS IN 5 EAST IS IN COMM LOSS. NIHON KOHDEN STAFF ON SITE AND DEEMED THAT THE SWITCHES WERE NON FUNCTIONAL AND OPENED A TICKET WITH (B)(6). NK STAFF WORKED ON GETTING THE CONFIGURATION DONE FOR THE SWITCHES AND HEADED BACK ON SITE. THEY REPLACED THE SWITCHES, AND EVERYTHING WAS BACK UP RUNNING. DEAD SWITCHES WILL BE SENT BACK TO JUNIPER FOR INVESTIGATION. THE CNS ITSELF WAS NOT RETURNED, BUT THE SWITCHES USED WITH THE NETWORK WERE RETURNED. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

THE NURSE REPORTED THAT THEY ARE EXPERIENCING A COMMUNICATION LOSS WITH A CENTRAL NURSE'S STATION (CNS) IN EAST AND 5 WEST. THEY STATED THERE IS CONSTRUCTION GOING ON WHICH KNOCKED OUT POWER TO NK DEVICES AND OTHER KAISER OWNED EQUIPMENT EARLIER IN THE DAY.

Description of Event or Problem · 0

THE NURSE REPORTED THAT THEY ARE EXPERIENCING A COMMUNICATION LOSS WITH A CENTRAL NURSE'S STATION (CNS) IN EAST AND 5 WEST. THEY STATED THERE IS CONSTRUCTION GOING ON WHICH KNOCKED OUT POWER TO NK DEVICES AND OTHER (B)(6) OWNED EQUIPMENT EARLIER IN THE DAY. POWER WAS RESTORED, BUT CNS IN 5 EAST IS IN COMM LOSS. NIHON KOHDEN STAFF ON SITE AND DEEMED THAT THE SWITCHES WERE NON FUNCTIONAL AND OPENED A TICKET WITH (B)(6). NK STAFF WORKED ON GETTING THE CONFIGURATION DONE FOR THE SWITCHES AND HEADED BACK ON SITE. THEY REPLACED THE SWITCHES, AND EVERYTHING WAS BACK UP RUNNING. DEAD SWITCHES WILL BE SENT BACK TO JUNIPER FOR INVESTIGATION. THE CNS ITSELF WAS NOT RETURNED, BUT THE SWITCHES USED WITH THE NETWORK WERE RETURNED. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
892742 CNS-6801A CENTRAL MONITOR SYSTEM MHX NIHON KOHDEN CORPORATION CNS-6801A NA 04931921131640

Patients

Seq Age Sex Outcome Treatment
1 Unknown TELEMETRY TRANSMITTER(S)| TELEMETRY TRANSMITTER(S)