FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 1199623
·
Received October 20, 2008
Report
- Report Number
- 8010493-2008-00020
- Event Type
- Injury
- Date Received
- October 20, 2008
- Date of Event
- September 16, 2008
- Report Date
- September 18, 2006
- Manufacturer
- KALTENBACH & VOIGT GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE PATIENT WAS GIVEN A TOPICAL ANAESTHETIC GEL IN THE OFFICE AND CONTINUED TO USE PARIOGUARD RINSE AT HOME. THE DOCTOR STATED THAT THE PATIENT HAS HEALED FINE.AT THE HANDPIECE EVALUATION, IT WAS FOUND THAT THE HEAD WAS DENTED AND THE BACKCAP WAS STUCK IN, CAUSING THE HEAD OF THE HANDPIECE TO HEAT UP.A REPLACEMENT HANDPIECE WAS SENT TO THE DOCTOR'S OFFICE.
Description of Event or Problem · 1
WHILE THE DENTIST WAS WORKING ON THE LOWER RIGHT TEETH, THE PATIENT'S CHEEK WAS BURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECES AND ACCESSORIES | EFB | KALTENBACH & VOIGT GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |