FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25LPA

MDR report key: 1199623 · Received October 20, 2008

Report

Report Number
8010493-2008-00020
Event Type
Injury
Date Received
October 20, 2008
Date of Event
September 16, 2008
Report Date
September 18, 2006
Manufacturer
KALTENBACH & VOIGT GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT WAS GIVEN A TOPICAL ANAESTHETIC GEL IN THE OFFICE AND CONTINUED TO USE PARIOGUARD RINSE AT HOME. THE DOCTOR STATED THAT THE PATIENT HAS HEALED FINE.AT THE HANDPIECE EVALUATION, IT WAS FOUND THAT THE HEAD WAS DENTED AND THE BACKCAP WAS STUCK IN, CAUSING THE HEAD OF THE HANDPIECE TO HEAT UP.A REPLACEMENT HANDPIECE WAS SENT TO THE DOCTOR'S OFFICE.

Description of Event or Problem · 1

WHILE THE DENTIST WAS WORKING ON THE LOWER RIGHT TEETH, THE PATIENT'S CHEEK WAS BURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25LPA DENTAL HANDPIECES AND ACCESSORIES EFB KALTENBACH & VOIGT GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention