FDA Adverse Event Malfunction Summary report: N

TX2 TISSUE REMOVAL SYSTEM - MICROTIP

MDR report key: 11995422 · Received June 14, 2021

Report

Report Number
1000135560-2021-00042
Event Type
Malfunction
Date Received
June 14, 2021
Date of Event
May 24, 2021
Report Date
June 14, 2021
Manufacturer
TENEX HEALTH, INC.
Product Code
LFL
UDI-DI
00857156006178
PMA / PMN Number
K153299
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND EVALUATED. IT WAS CONFIRMED THAT THE MICROTIP NEEDLE HAD SEPARATED FROM THE REST OF THE HANDPIECE. THE FRACTURE SITE DID NOT IMMEDIATELY INDICATE A SPECIFIC CAUSE FOR THE FAILURE. TESTING AND DISASSEMBLY OF THE DEVICE DID NOT REVEAL ANY FURTHER ANOMALIES OR DEFECTS.

Description of Event or Problem · 1

DURING A PROCEDURE WITH THE TENEX SYSTEM, A PORTION OF THE MICROTIP NEEDLE SEPARATED FROM THE REST OF THE HAND PIECE. THE PROCEDURE WAS COMPLETED WITH ANOTHER HAND PIECE. THERE WERE NO PATIENT COMPLICATIONS.