COULTER LH 750 ANALYZER
Report
- Report Number
- 1061932-2008-00052
- Event Type
- Other
- Date Received
- October 8, 2008
- Date of Event
- August 22, 2008
- Report Date
- October 8, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
SAMPLES WERE COLLECTED IN 4ML BD EDTA VACUTAINERS AND STORED AT ROOM TEMP UNTIL DRAWN (0 TO 4 HRS) LATER. PREVIOUSLY-RUN PT SAMPLES WERE RERUN TO CONFIRM BACK TO THE LAST ACCEPTABLE CONTROL RUN. QC IS RUN 3 TIMES PER DAY. QC WAS RUN BEFORE AND AFTER THE INCIDENT AND RECOVERED WITHIN ACCEPTABLE RANGES. THE UNIT IS CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS. TWO PIECES OF TUBING ATTACHED TO THE NEEDLE VENT CHAMBER - A VACUUM/PRESSURE SUPPLY AND THE NEEDLE VENT LINE WERE REVERSED AS TO WHICH PORT THEY WERE CONNECTED, CAUSING THE NEEDLE VENT LINE TO DRAW FLUID FROM THE BOTTOM OF THE NEEDLE VENT CHAMBER BACK INTO THE PT SAMPLE. FIELD SERVICE ENGINEER (FSE) WAS ON-SITE IN 2008 AND CORRECTLY RETUBED THE NEEDLE VENT CHAMBER. THE IMPROPERLY REVERSED TUBING ON THE NEEDLE VENT CHAMBER MAY HAVE CONTRIBUTED TO THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUS HEMOGLOBIN (HGB) AND PLATELETS (PLT) RESULTS GENERATED BY THE COULTER LH 750 ANALYZER FOR SEVERAL PTS. THE PT SAMPLE HGB RESULTS PROVIDED BY THE CUSTOMER WERE IN THE RANGE OF 6.1 - 7.5G/DL. UPON RE-TESTING ON A DIFFERENT INSTRUMENT, THE RESULTS WERE IN THE RANGE OF 5.2 - 7.6G/DL THE PLT RESULTS PROVIDED BY THE CUSTOMER WERE IN THE RANGE 112 - 272 X 10 TO THE THIRD POWER CELLS/UL AND WHEN TESTED ON A DIFFERENT INSTRUMENT, THE RESULTS WERE IN THE RANGE OF 113 - 259 X 10 TO THE THIRD POWER CELLS/UL. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LAB AND QUESTIONED BY THE PHYSICIANS. RESULTS WERE BELIEVED TO BE ERRONEOUS BECAUSE HEMATOCRIT (HCT) RESULTS SIGNIFICANTLY CHANGED FROM PREVIOUS PT RESULTS. ONE PT WAS TRANSFUSED AND ONE PT WAS HELD BACK AN ADD'L 24 HRS IN ICU AND AT LEAST THREE PTS THAT WERE REFERRED TO THE HOSP TO DETERMINE IF TRANSFUSIONS WERE NEEDED BUT THEY WERE REDRAWN, RETESTED, AND WERE NOT TRANSFUSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER LH 750 ANALYZER | HEMATOLOGY ANALYZER | GKZ | BECKMAN COULTER, INC. | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |