FDA Adverse Event Injury Summary report: N

PENTAX

MDR report key: 11988815 · Received June 12, 2021

Report

Report Number
2518897-2021-00131
Event Type
Injury
Date Received
June 12, 2021
Date of Event
May 13, 2021
Report Date
June 2, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDS
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
KS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SUMMARY: AFTER RETURNING THIS ENDOSCOPE, IT WAS INSPECTED AND THE OTHER TWO SCREWS WERE FOUND IN PMK BY THE SERVICE TEAM. PENTAX R&D TEAM HAS STARTED INVESTIGATION FOR THIS ISSUE(BENDING RUBBER PEEL OFF AND LOOSED SCREWS) BUT ROOT CAUSE IS NOT DETERMINED AS OF NOW BECAUSE RE-CREATION IS IMPOSSIBLE THROUGH THEIR ACTIVITIES. THEREFORE WE COULD PROVIDE NO SPECIAL RECOMMENDATION TO AVOID/MITIGATE FAILURE. WEAR WAS SELECTED AS THE MOST APPROPRIATE INVESTIGATION CODE. REPLACING BELOW SPARE PART: Q010-U5020 SEGMENT STAYCOIL ASSY 1 AND D965-SA038 BENDING RUBBER 1. PENTAX MEDICAL HAS NOT RECEIVED ANY FURTHER INFORMATION FOR THIS EVENT AND THEREFORE, CONSIDERS THIS MEDWATCH REPORT CLOSED.

Description of Event or Problem · 1

PENTAX MEDICAL WAS INFORMED OF AN EVENT ON (B)(6) 2021 THAT OCCURRED IN SOUTH KOREA WITHIN THE APAC REGION DURING USE. THE INFORMATION PROVIDED INDICATED THAT DURING THE ENDOSCOPY THE BENDING RUBBER BURST AND A SCREW FELL OUT. THE ENDOSCOPE WAS WITHDRAWN RIGHT AWAY, AND ANOTHER ENDOSCOPE WAS INSERTED TO REMOVE THE SCREW THAT FELL OUT INVOLVING PENTAX MEDICAL VIDEO GASTROSCOPE, MODEL EG29-I10C, SERIAL NUMBER (B)(4). AFTER RETURNING THIS ENDOSCOPE, IT WAS INSPECTED AND THE OTHER TWO SCREWS WERE FOUND OUT IN PMK BY THE SERVICE TEAM. CURRENTLY THE ENDOSCOPE IS AWAITING REPAIR. THE INVESTIGATION IS IN-PROCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
885111 PENTAX VIDEO GASTROSCOPE FDS HOYA CORPORATION PENTAX TOKYO OFFICE EG29-I10C

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention