FDA Adverse Event Malfunction Summary report: N

PN RELION 31G X 8MM 3B TW

MDR report key: 11981042 · Received June 11, 2021

Report

Report Number
3014704491-2021-00022
Event Type
Malfunction
Date Received
June 11, 2021
Date of Event
May 18, 2021
Report Date
June 21, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FMI
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. CLOG TEST WAS CONDUCTED ON 7PCS RETENTION SAMPLES AND NO NEEDLE CLOG FAIL FOUND. BASED ON INVESTIGATION THE CERTAIN CAUSE CANNOT BE CONCLUDED AT THE MOMENT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 PN RELION 31G X 8MM 3B TW NEEDLE WAS UNABLE TO DELIVER INSULIN OR MEDICATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER :   THE CONSUMER REPORTED THAT THE NEEDLES DO NOT WORK AND ARE CLOGGED. DATE OF EVENT: UNKNOWN. SAMPLES: DISCARDED.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS PENDING BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 1 PN RELION 31G X 8MM 3B TW NEEDLE WAS UNABLE TO DELIVER INSULIN OR MEDICATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER :   THE CONSUMER REPORTED THAT THE NEEDLES DO NOT WORK AND ARE CLOGGED. DATE OF EVENT: UNKNOWN, SAMPLES: DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
881938 PN RELION 31G X 8MM 3B TW HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 0140342

Patients

Seq Age Sex Outcome Treatment
1