FDA Adverse Event Injury Summary report: N

PRECISION MONTAGE MRI

MDR report key: 11980132 · Received June 10, 2021

Report

Report Number
3006630150-2021-02843
Event Type
Injury
Date Received
June 10, 2021
Date of Event
April 26, 2021
Report Date
January 31, 2022
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
UDI-DI
08714729905943
PMA / PMN Number
P030017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317700. MODEL: SC-2317-70. SERIAL: (B)(6). BATCH: 7072232.

Additional Manufacturer Narrative · 0

SC-1200, SERIAL (B)(6): ENGINEERS INSPECTED AND ANALYZED THIS DEVICE UPON RECEIPT. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM RESET COUNT WAS 3, THAT SUGGESTED NO STIMULATION INTERRUPTION DUE TO IPG MALFUNCTION PRIOR TO THE PRODUCT RETURN. MONITORED STIMULATION ON AN OSCILLOSCOPE FOUND THE OUTPUTS WERE CONSISTENT AND CORRECT ON ALL ELECTRODES. CURRENT LEAKAGE TESTS AND RESIDUAL GAS ANALYSIS VERIFIED NO LOSS OF ELECTRIC CURRENT INTO THE SURROUNDING TISSUE. THE COMPLAINT OF THE IPG SWITCHING OFF WAS NOT CONFIRMED. IT PASSED THE FUNCTIONAL TEST, AND AN ELECTRICAL TEST REVEALED NORMAL DEVICE CHARACTERISTICS. SC-2317-70, SERIAL (B)(6): THE COMPLAINT OF HIGH IMPEDANCES ON ONE OF THE LEADS WAS CONFIRMED. VISUAL AND X-RAY INSPECTIONS OF THE RETURNED LEAD CONFIRMED SEVEN CABLES WERE FRACTURED AT THE CLIK SITE, ABOUT 15.5 CENTIMETERS FROM THE DISTAL END THAT RESULTED IN THE SOURCE OF HIGH IMPEDANCES.

Additional Manufacturer Narrative · 0

SC-1200, (B)(6). ENGINEERS INSPECTED AND ANALYZED THIS DEVICE UPON RECEIPT. THE IPG SYSTEM RESET COUNT WAS 3, THAT SUGGESTED NO STIMULATION INTERRUPTION DUE TO IPG MALFUNCTION PRIOR TO THE PRODUCT RETURN. MONITORED STIMULATION ON AN OSCILLOSCOPE FOUND THE OUTPUTS WERE CONSISTENT AND CORRECT ON ALL ELECTRODES. CURRENT LEAKAGE TESTS AND RESIDUAL GAS ANALYSIS VERIFIED NO LOSS OF ELECTRIC CURRENT INTO THE SURROUNDING TISSUE. THE HIGHEST TEMPERATURE DURING CHARGING CYCLES IN THE PATIENTS BODY REGISTERED WITHIN THE ACCEPTABLE RANGE IT PASSED THE FUNCTIONAL TEST, AND AN ELECTRICAL TEST REVEALED NORMAL DEVICE CHARACTERISTICS. SC-2317-70, (B)(6). THE COMPLAINT OF HIGH IMPEDANCES ON ONE OF THE LEADS WAS CONFIRMED. VISUAL AND X-RAY INSPECTIONS OF THE RETURNED LEAD CONFIRMED SEVEN CABLES WERE FRACTURED AT THE CLIK SITE, ABOUT 15.5 CENTIMETERS FROM THE DISTAL END THAT RESULTED IN THE SOURCE OF HIGH IMPEDANCES. WHEN EXCESSIVE MECHANICAL-TENSILE FORCE WAS EXERTED ONTO THE LEAD, THE LEAD BECAME KINKED AFTER EXITING THE CLIK ANCHOR RESULTING IN THE CABLE FRACTURES.

Additional Manufacturer Narrative · 0

SC-1200, SERIAL (B)(6): ENGINEERS INSPECTED AND ANALYZED THIS DEVICE UPON RECEIPT. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM RESET COUNT WAS 3, THAT SUGGESTED NO STIMULATION INTERRUPTION DUE TO IPG MALFUNCTION PRIOR TO THE PRODUCT RETURN. MONITORED STIMULATION ON AN OSCILLOSCOPE FOUND THE OUTPUTS WERE CONSISTENT AND CORRECT ON ALL ELECTRODES. CURRENT LEAKAGE TESTS AND RESIDUAL GAS ANALYSIS VERIFIED NO LOSS OF ELECTRIC CURRENT INTO THE SURROUNDING TISSUE. THE COMPLAINT OF THE IPG SWITCHING OFF WAS NOT CONFIRMED. IT PASSED THE FUNCTIONAL TEST, AND AN ELECTRICAL TEST REVEALED NORMAL DEVICE CHARACTERISTICS. SC-2317-70, SERIAL (B)(6): THE COMPLAINT OF HIGH IMPEDANCES ON ONE OF THE LEADS WAS CONFIRMED. VISUAL AND X-RAY INSPECTIONS OF THE RETURNED LEAD CONFIRMED SEVEN CABLES WERE FRACTURED AT THE CLIK SITE, ABOUT 15.5 CENTIMETERS FROM THE DISTAL END THAT RESULTED IN THE SOURCE OF HIGH IMPEDANCES.

Description of Event or Problem · 0

IT WAS REPORTED THAT APPROXIMATELY SIX MONTHS AFTER THE PERMANENT IMPLANT PROCEDURE THE IMPLANTABLE PULSE GENERATOR (IPG) STARTED TO SWITCH OFF ON ITS OWN. IN ADDITION, THE PATIENT WAS EXPERIENCING A PARTIAL LOSS OF THERAPY. AN IMPEDANCE CHECK REVEALED MULTIPLE CONTACTS OF THE LEAD DISPLAYED HIGH IMPEDANCES. REPROGRAMMING WAS ATTEMPTED, HOWEVER THE IPG CONTINUED SWITCHING OFF. THE PATIENT REPORTED THERE WERE NO SITUATIONS THAT WOULD HAVE EXPOSED THE PATIENT TO LARGE MAGNETIC FIELDS ETC. THE PATIENT UNDERWENT A REVISION PROCEDURE IN WHICH THE PHYSICIAN REPLACED THE ENTIRE SYSTEM. THE EXPLANTED IPG AND LEAD WILL BE RETURNED. THE PATIENT IS DOING WELL POST OPERATIVELY.

Description of Event or Problem · 0

IT WAS REPORTED THAT APPROXIMATELY SIX MONTHS AFTER THE PERMANENT IMPLANT PROCEDURE THE IMPLANTABLE PULSE GENERATOR (IPG) STARTED TO SWITCH OFF ON ITS OWN. IN ADDITION, THE PATIENT WAS EXPERIENCING A PARTIAL LOSS OF THERAPY. AN IMPEDANCE CHECK REVEALED MULTIPLE CONTACTS OF THE LEAD DISPLAYED HIGH IMPEDANCES. REPROGRAMMING WAS ATTEMPTED, HOWEVER THE IPG CONTINUED SWITCHING OFF. THE PATIENT REPORTED THERE WERE NO SITUATIONS THAT WOULD HAVE EXPOSED THE PATIENT TO LARGE MAGNETIC FIELDS ETC. THE PATIENT UNDERWENT A REVISION PROCEDURE IN WHICH THE PHYSICIAN REPLACED THE ENTIRE SYSTEM. THE EXPLANTED IPG AND LEAD WILL BE RETURNED. THE PATIENT IS DOING WELL POST OPERATIVELY. ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS ALSO EXPERIENCING A WARMING SENSATION AT THE IPG SITE.

Description of Event or Problem · 0

IT WAS REPORTED THAT APPROXIMATELY SIX MONTHS AFTER THE PERMANENT IMPLANT PROCEDURE THE IMPLANTABLE PULSE GENERATOR (IPG) STARTED TO SWITCH OFF ON ITS OWN. IN ADDITION, THE PATIENT WAS EXPERIENCING A PARTIAL LOSS OF THERAPY. AN IMPEDANCE CHECK REVEALED MULTIPLE CONTACTS OF THE LEAD DISPLAYED HIGH IMPEDANCES. REPROGRAMMING WAS ATTEMPTED, HOWEVER THE IPG CONTINUED SWITCHING OFF. THE PATIENT REPORTED THERE WERE NO SITUATIONS THAT WOULD HAVE EXPOSED THE PATIENT TO LARGE MAGNETIC FIELDS ETC. THE PATIENT UNDERWENT A REVISION PROCEDURE IN WHICH THE PHYSICIAN REPLACED THE ENTIRE SYSTEM. THE EXPLANTED IPG AND LEAD WILL BE RETURNED. THE PATIENT IS DOING WELL POST OPERATIVELY. ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS ALSO EXPERIENCING A WARMING SENSATION AT THE IPG SITE.

Description of Event or Problem · 0

IT WAS REPORTED THAT APPROXIMATELY SIX MONTHS AFTER THE PERMANENT IMPLANT PROCEDURE THE IMPLANTABLE PULSE GENERATOR (IPG) STARTED TO SWITCH OFF ON ITS OWN. IN ADDITION, THE PATIENT WAS EXPERIENCING A PARTIAL LOSS OF THERAPY. AN IMPEDANCE CHECK REVEALED MULTIPLE CONTACTS OF THE LEAD DISPLAYED HIGH IMPEDANCES. REPROGRAMMING WAS ATTEMPTED, HOWEVER THE IPG CONTINUED SWITCHING OFF. THE PATIENT REPORTED THERE WERE NO SITUATIONS THAT WOULD HAVE EXPOSED THE PATIENT TO LARGE MAGNETIC FIELDS ETC. THE PATIENT UNDERWENT A REVISION PROCEDURE IN WHICH THE PHYSICIAN REPLACED THE ENTIRE SYSTEM. THE EXPLANTED IPG AND LEAD WILL BE RETURNED. THE PATIENT IS DOING WELL POST OPERATIVELY.

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317700. MODEL: SC-2317-70. SERIAL: (B)(4). BATCH: 7072232.

Description of Event or Problem · 1

IT WAS REPORTED THAT APPROXIMATELY SIX MONTHS AFTER THE PERMANENT IMPLANT PROCEDURE THE IMPLANTABLE PULSE GENERATOR (IPG) STARTED TO SWITCH OFF ON ITS OWN. IN ADDITION, THE PATIENT WAS EXPERIENCING A PARTIAL LOSS OF THERAPY. AN IMPEDANCE CHECK REVEALED MULTIPLE CONTACTS OF THE LEAD DISPLAYED HIGH IMPEDANCES. REPROGRAMMING WAS ATTEMPTED, HOWEVER THE IPG CONTINUED SWITCHING OFF. THE PATIENT REPORTED THERE WERE NO SITUATIONS THAT WOULD HAVE EXPOSED THE PATIENT TO LARGE MAGNETIC FIELDS ETC. THE PATIENT UNDERWENT A REVISION PROCEDURE IN WHICH THE PHYSICIAN REPLACED THE ENTIRE SYSTEM. THE EXPLANTED IPG AND LEAD WILL BE RETURNED. THE PATIENT IS DOING WELL POST OPERATIVELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
873667 PRECISION MONTAGE MRI STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1200 370617 08714729905943

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention