FDA Adverse Event Death Summary report: N

EDWARDS TRANSFEMORAL ESHEATH

MDR report key: 11979807 · Received June 10, 2021

Report

Report Number
2015691-2021-03483
Event Type
Death
Date Received
June 10, 2021
Date of Event
January 1, 2021
Report Date
June 10, 2021
Manufacturer
EDWARDS LIFESCIENCES LLC
Product Code
NPT
PMA / PMN Number
P140031
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016006. THIS REPORT PROVIDES DATA FROM (B)(6) REGISTRY EXEMPTION NUMBER E2016006 AND SUMMARIZES 3 EVENTS OF RETROPERITONEAL BLEEDING DEATH EVENTS FOR THE SAPIEN 3 ULTRA / EDWARDS AXELA SHEATH / EDWARDS ESHEATH INTRODUCER SET IN THE AORTIC POSITION. THE 'TIME TO EVENT' (TTE, IN DAYS) FOR THIS EVENT WAS 0.00. DUE TO THE LIMITED INFORMATION AND THE DATA RECEIVED, EDWARDS CANNOT CONFIRM WHICH ESHEATH WAS USED IN THIS CASE, THE AXELA OR ESHEATH INTRODUCER SET. THE UDI FOR BOTH DEVICES ARE REFERENCED BELOW. THE DEVICE IDENTIFICATION (DI) NUMBERS FOR EDWARDS AXELA SHEATH IS (B)(4). THE DEVICE IDENTIFICATION (DI) NUMBERS FOR EDWARDS ESHEATH INTRODUCER SET ARE (B)(4). THE INSTRUCTIONS FOR USE (IFU) LIST HEMORRHAGE REQUIRING TRANSFUSION OR INTERVENTION AS A POTENTIAL RISK ASSOCIATED WITH THE OVERALL TAVR PROCEDURE, INCLUDING POTENTIAL ACCESS COMPLICATIONS ASSOCIATED WITH STANDARD CARDIAC CATHETERIZATION. THE MAJORITY OF TRANSFEMORAL ACCESS RELATED BLEEDING COMPLICATIONS ARE RELATED TO DISEASED ILEO-FEMORAL VESSELS AND/OR PROCEDURAL TECHNIQUE DURING THE INSERTION OR REMOVAL OF THE SHEATH AND/OR DILATORS. THERE ARE CASES WHERE THE BLEEDING IS SIGNIFICANT AND MORE COMPLEX INTERVENTION OR TRANSFUSION IS REQUIRED TO TREAT/PREVENT A PERMANENT INJURY FROM OCCURRING. THE THV PHYSICIAN TRAINING MANUALS INSTRUCT ON PROCEDURAL CONSIDERATIONS FOR SHEATH INSERTION WITH REGARDS TO PROPER SCREENING CRITICAL TO REDUCING VASCULAR COMPLICATIONS. THE IFU CONTRAINDICATES PATIENTS WITH ILIO-FEMORAL VESSEL CHARACTERISTICS THAT WOULD PRECLUDE SAFE PLACEMENT OF SHEATHS SUCH AS SEVERE OBSTRUCTIVE CALCIFICATION OR SEVERE TORTUOSITY. PRE-PROCEDURE SCREENING AND ASSESSMENT OF THE FEMORAL/ILIAC ARTERY INTERNAL DIAMETERS WILL ENABLE THE CLINICIAN TO DETERMINE IF THE VALVE CAN BE DELIVERED TRANSFEMORALLY. IN THIS CASE, SPECIFIC PROCEDURAL DETAILS ARE NOT AVAILABLE TO DETERMINE POTENTIAL CONTRIBUTING FACTORS TO THE EVENT, OR IF THE EVENT IS RELATED TO AN EDWARDS DEVICE; HOWEVER, A TRANSFUSION OF 2 UNITS OR MORE OF BLOOD WAS REPORTED IN THE REGISTRY. BY VARC DEFINITION, THE MEET CRITERIA FOR MAJOR VASCULAR COMPLICATION THEREFORE, THE EVENT IS BEING REPORTED. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME. SUMMARY REPORT.

Description of Event or Problem · 1

(B)(6) REGISTRY SUMMARY REPORTING FOR ADVERSE EVENT SUBMISSION FOR QUARTER 1 2021 DATA EXTRACT FOR AORTIC DEATHS FOR THE SAPIEN 3 ULTRA VALVE. THIS REPORT SUMMARIZES 3 RETROPERITONEAL BLEEDING DEATH EVENTS FOR THE SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE / EDWARDS AXELA SHEATH / EDWARDS ESHEATH INTRODUCER SET. THE AGE RANGE FOR THESE EVENTS IS (B)(6) YEARS. THE BREAKDOWN FOR GENDER IS AS FOLLOWS: 2 FEMALES AND 1 MALE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
874609 EDWARDS TRANSFEMORAL ESHEATH AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES LLC TF SHEATH UNKNOWN N/A

Patients

Seq Age Sex Outcome Treatment
1 85 YR Death