INFUSOMAT
Report
- Report Number
- 9610825-2021-00228
- Event Type
- Malfunction
- Date Received
- June 10, 2021
- Date of Event
- May 21, 2021
- Report Date
- August 5, 2021
- Manufacturer
- B.BRAUN MELSUNGEN AG
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4). THE DEVICE WAS INVESTIGATED IN OUR LABORATORY IN SAO GONZALO, BRAZIL: ON (B)(6) 2021 THE FLOW RATE OF 288ML/H WAS NOT PROGRAMMED. BUT WE FOUND A SCHEDULE WITH AN APPROXIMATE VALUE OF 280ML/H, AT 15:48:56, FOR A VOLUME OF 560ML, TO OCCUR IN 02 HOURS. THE HISTORY SHOWS THAT THE SCHEDULE HAPPENED EXACTLY AS PER THE SCHEDULE". "AS THE CUSTOMER REPORTED THAT THE EQUIPMENT DID NOT INFUSE THE DRUG, WE SIMULATED AN INFUSION USING THE SAME PARAMETERS FOUND IN THE USAGE HISTORY AND THE EQUIPMENT WORKED PERFECTLY AS EXPECTED.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4).THE COMPLAINT IS UNDER EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN (B)(6): "UNDERINFUSION" ACCORDING TO THE CUSTOMER: "THE CUSTOMER REPORTS THAT THE PUMP WAS PROGRAMMED FOR 2 HOURS, WITH OXALIPLATIN MEDICINE, WITH A FLOW RATE OF 288ML/H, BUT THE MEDICATION HAS NOT BEEN INFUSED. THE CUSTOMER INFORMS THAT THIS FACT OCCURRED ON (B)(6) 2021."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 870929 | INFUSOMAT | PUMP, INFUSION | FRN | B.BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |