FDA Adverse Event Malfunction Summary report: N

S5 IMAGING SYSTEM

MDR report key: 1197623 · Received October 13, 2008

Report

Report Number
2939520-2008-00023
Event Type
Malfunction
Date Received
October 13, 2008
Date of Event
September 16, 2008
Report Date
October 14, 2008
Manufacturer
VOLCANO CORP
Product Code
IYO
PMA / PMN Number
K061215
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM AND MATERIALS HAVE BEEN DESIGNED AND TESTED TO IEC 60601-1 TO ENSURE THAT THE POSSIBILITY OF A FIRE OCCURRING IS VERY REMOTE. ALTHOUGH, THERE WAS NO INJURY THAT OCCURRED DUE TO THIS MALFUNCTION, THERE IS THE POSSIBILITY THAT IF IT WERE TO RECUR, A NEGATIVE IMPACT TO OTHER EQUIPMENT AND/OR MATERIALS IN THE ROOM (I.E. COMPRESSED GAS TANKS, FLAMMABLE MATERIALS, ECT.) COULD RESULT IN HARM TO THE PT/USER. THIS REPORT IS BEING SENT AS A NOTIFICATION.

Description of Event or Problem · 1

SMOKE CAME OUT OF THE BACK OF THE SYSTEM WHEN IT WAS TURNED ON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S5 IMAGING SYSTEM IVUS IMAGING SYSTEM IYO VOLCANO CORP 804200001-R

Patients

Seq Age Sex Outcome Treatment
1