FDA Adverse Event Malfunction Summary report: N

NEOTECH PRODUCTS

MDR report key: 11974562 · Received June 10, 2021

Report

Report Number
2025917-2021-00005
Event Type
Malfunction
Date Received
June 10, 2021
Report Date
June 10, 2021
Product Code
BXJ
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS NOT RETURNED AND NO EVALUATION WAS POSSIBLE ON THE ACTUAL DEVICE. HOWEVER, SAMPLES OF FINISHED GOODS WERE PULLED OUT FROM RETAINED SAMPLES OF BOTH REPORTED AND CURRENT LOT NUMBERS FOR RE-INSPECTION AND NO ABNORMALITY WAS FOUND. ALSO, THE PRODUCT'S DEVICE HISTORY RECORDS (DHR) OF REPORTED LOT AND CURRENT INVENTORY WERE REVIEWED AND NO NONCONFORMITY WAS FOUND. PER OUR INVESTIGATION DEVICE WAS USED WITH RAM CANNULA AND ON OFF LABEL USE FOR CPAP. THIS COULD BE CONTRIBUTED TO THESE ISOLATED INCIDENTS AS WE JUST HAD ONE SIMILAR INCIDENT FROM THE SAME SOURCE IN THE PAST FIVE YEARS FOR MILLIONS OF THE PARTS THAT WERE USED ALL AROUND THE WORLD. PER COMPLAINTS RATIO, THESE INCIDENTS WERE ISOLATED EVENTS. THIS COMPLAINT WILL BE MONITORED FOR TRENDING PURPOSES AND IS ADDED TO THE RELATED LOGS AND CHARTS.

Description of Event or Problem · 1

PER CUSTOMER, NEOSEAL FOUND IN BABY'S MOUTH BUT WAS REMOVED IN TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
869068 NEOTECH PRODUCTS NEOSEAL BXJ N422 2020-0813

Patients

Seq Age Sex Outcome Treatment
1 Other