FDA Adverse Event Malfunction Summary report: N

BD SAF-T-INTIMA SAFETY SYSTEM WITH Y ADAPTER 24 GA 0.75 IN

MDR report key: 11971419 · Received June 9, 2021

Report

Report Number
9610847-2021-00265
Event Type
Malfunction
Date Received
June 9, 2021
Date of Event
February 8, 2021
Report Date
June 18, 2021
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FOZ
UDI-DI
30382903833130
PMA / PMN Number
K923702
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED LOT NUMBER 9252066. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH Y ADAPTER 24 GA 0.75 IN HAD THE SAFETY MECHANISM FAIL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED BY THE FACILITY REPRESENTATIVE THAT THE NEEDLE DID NOT RETRACT INTO THE SAFETY SHIELD."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH Y ADAPTER 24 GA 0.75 IN HAD THE SAFETY MECHANISM FAIL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED BY THE FACILITY REPRESENTATIVE THAT THE NEEDLE DID NOT RETRACT INTO THE SAFETY SHIELD."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
867719 BD SAF-T-INTIMA SAFETY SYSTEM WITH Y ADAPTER 24 GA 0.75 IN INTRAVASCULAR CATHETER FOZ BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 383313 9252066 30382903833130

Patients

Seq Age Sex Outcome Treatment
1