FDA Adverse Event Death Summary report: Y

EDWARDS COMMANDER DELIVERY SYSTEM

MDR report key: 11971375 · Received June 9, 2021

Report

Report Number
2015691-2021-03393
Event Type
Death
Date Received
June 9, 2021
Date of Event
January 1, 2021
Report Date
June 9, 2021
Manufacturer
EDWARDS LIFESCIENCES LLC
Product Code
NPT
PMA / PMN Number
P140031
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT PROVIDES DATA FROM THV/TVT REGISTRY DATA EXEMPTION NUMBER E2016006 AND SUMMARIZES 10 EVENTS OF PERFORATION WITH OR W/O TAMPONADE FOR THE SAPIEN 3 VALVE IN THE AORTIC POSITION. THE TIME TO EVENT [TTE IN DAYS] FOR THIS EVENT WAS 0.2. THE DEVICE IDENTIFICATION (DI) NUMBERS FOR EDWARDS COMMANDER DELIVERY SYSTEM ARE: (B)(4). PER THE INSTRUCTIONS FOR USE, CARDIOVASCULAR INJURIES, INCLUDING PERFORATION OR DISSECTION OF VESSELS, VENTRICLE, MYOCARDIUM OR VALVULAR STRUCTURES, ARE POTENTIAL ADVERSE EVENTS ASSOCIATED WITH STANDARD CARDIAC CATHETERIZATION, BALLOON VALVULOPLASTY AND THE TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE. THERE ARE SEVERAL POTENTIAL ETIOLOGIES FOR VENTRICULAR PERFORATION DURING A TAVR PROCEDURE, INCLUDING PERFORATION BY THE GUIDEWIRE, THE DELIVERY SYSTEM, OR THE TRANSVENOUS PACER (TVP) LEAD. PHYSICIANS ARE EXTENSIVELY TRAINED BY EDWARDS BEFORE THEY ARE QUALIFIED TO USE THE TRANSCATHETER HEART VALVE (THV). TRAINING INCLUDES PROPER GUIDEWIRE POSITIONING, FIXATION OF THE TVP TO PREVENT VENTRICLE PERFORATION, AND CAREFUL MANIPULATION OF DEVICES. PER THE PROCEDURE DIDACTIC, PATIENTS WITH SMALL VENTRICLES ARE AT PARTICULARLY HIGH RISK FOR VENTRICULAR PERFORATION. IN THIS CASE, SPECIFIC PROCEDURAL DETAILS ARE NOT AVAILABLE TO DETERMINE POTENTIAL CONTRIBUTING FACTORS TO THE EVENT. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THV/TVT REGISTRY SUMMARY REPORTING FOR ADVERSE EVENT SUBMISSION FOR Q1 2021 DATA EXTRACT FOR AORTIC DEATH FOR THE SAPIEN 3 VALVE. THIS REPORT SUMMARIZES 10 PERFORATION WITH OR W/O TAMPONADE EVENTS FOR THE SAPIEN 3 VALVE. THE AGE RANGE FOR THESE EVENTS IS FROM 55 TO 93. THE BREAKDOWN FOR GENDER IS AS FOLLOWS: 7 MALES AND 3 FEMALES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
867429 EDWARDS COMMANDER DELIVERY SYSTEM AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES LLC COMMANDER DELIVERY SYSTEM UNKNOWN N/A

Patients

Seq Age Sex Outcome Treatment
1 80 YR Death