Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 221270, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 1092958 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE PHYSICAL ATTRIBUTE TESTING. A SAMPLE OF PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. THEY ARE TESTED FOR PHYSICAL ATTRIBUTES PRIOR TO RELEASE TO ENSURE THAT THEY CONFORM TO PRODUCT SPECIFICATIONS. ALL PHYSICAL ATTRIBUTE TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER BD INTERNAL PROCEDURES. THE COMPLAINT HISTORY WAS REVIEWED, AND ONE OTHER COMPLAINT HAS BEEN TAKEN ON BATCH 1092958 FOR CONTAMINATION AND SPLIT AGAR. RETENTION SAMPLES FROM BATCH 1092958 WERE NOT AVAILABLE FOR INSPECTION. ONE PHOTO WAS RECEIVED FOR INVESTIGATION. THE PHOTO SHOWS THE BOTTOM OF TWO PLATES FROM BATCH 1092858 (TIME STAMP 0634). ONE PLATE HAS A SPLIT IN THE AGAR BED AND THE OTHER PLATE HAS MICROBIAL GROWTH. NO RETURN SAMPLES WERE RECEIVED FOR INVESTIGATION. THIS COMPLAINT CAN BE CONFIRMED. BASED ON THE LOW DEFECT RATE FOR THIS BATCH; NO ACTIONS ARE PLANNED AT THIS TIME. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION AND AGAR DEFECTS. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING 4 BD BBL¿ MACCONKEY II AGAR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "REPORTED PLATES APPEAR TO BE CONTAMINATED WITH GROWTH. MEDIA 221270 LOT 1092958. "

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING 4 BD BBL¿ MACCONKEY II AGAR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "REPORTED PLATES APPEAR TO BE CONTAMINATED WITH GROWTH. MEDIA 221270 LOT 1092958. "

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
865645 BD BBL¿ MACCONKEY II AGAR CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL JSI BECTON, DICKINSON & CO. (SPARKS) 221270 1092958 10382902212703

Patients

Seq Age Sex Outcome Treatment
1 Unknown