FDA Adverse Event Other Summary report: N

ADIVA CHEMISTRY TOTAL PROTEIN REAGENT KIT

MDR report key: 1196535 · Received October 10, 2008

Report

Report Number
2432235-2008-00118
Event Type
Other
Date Received
October 10, 2008
Date of Event
September 12, 2008
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
CEK
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

REMEDIAL ACTION: THE DEXTRAN INTERFERENCE SECTION IN THE IFU WILL BE REVISED TO REFLECT CURRENT DATA.

Description of Event or Problem · 1

DURING STUDIES, IT WAS OBSERVED THAT THE ADVIA CHEMISTRY TOTAL PROTEIN METHOD DID NOT MEET CLAIMS OF NO SIGNIFICANT INTERFERENCE AT 1500 MG/DL DEXTRAN. THERE HAVE BEEN NO REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO POSITIVE DISCORDANT RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADIVA CHEMISTRY TOTAL PROTEIN REAGENT KIT TOTAL PROTEIN CHEMISTRY METHOD CEK SIEMENS HEALTHCARE DIAGNOSTICS INC. NA 100789

Patients

Seq Age Sex Outcome Treatment
1