FDA Adverse Event
Other
Summary report: N
ADIVA CHEMISTRY TOTAL PROTEIN REAGENT KIT
MDR report key: 1196535
·
Received October 10, 2008
Report
- Report Number
- 2432235-2008-00118
- Event Type
- Other
- Date Received
- October 10, 2008
- Date of Event
- September 12, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- CEK
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
REMEDIAL ACTION: THE DEXTRAN INTERFERENCE SECTION IN THE IFU WILL BE REVISED TO REFLECT CURRENT DATA.
Description of Event or Problem · 1
DURING STUDIES, IT WAS OBSERVED THAT THE ADVIA CHEMISTRY TOTAL PROTEIN METHOD DID NOT MEET CLAIMS OF NO SIGNIFICANT INTERFERENCE AT 1500 MG/DL DEXTRAN. THERE HAVE BEEN NO REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO POSITIVE DISCORDANT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADIVA CHEMISTRY TOTAL PROTEIN REAGENT KIT | TOTAL PROTEIN CHEMISTRY METHOD | CEK | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | 100789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |