FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11963178 · Received June 8, 2021

Report

Report Number
1221359-2021-01635
Event Type
Malfunction
Date Received
June 8, 2021
Date of Event
May 10, 2021
Report Date
August 11, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1022242 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1022242, TEST BASE PART NUMBER 190-430/ LOT: 1022242. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 152003 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Additional Manufacturer Narrative · 1

(B)(6). THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A NASAL RAYON SWAB. AS PER THE CUSTOMER REPEATING TESTING WAS PERFORMED ON THE ID NOW AND GENERATED NEGATIVE RESULTS. RT-PCR CONFIRMATION TESTING GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT HAD A FEVER. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
850446 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC 1022242 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 29 YR