FDA Adverse Event Malfunction Summary report: N

MOTOBAND DYNAFORCE DYNABUNION

MDR report key: 11961667 · Received June 8, 2021

Report

Report Number
3011421599-2021-00010
Event Type
Malfunction
Date Received
June 8, 2021
Date of Event
March 24, 2021
Report Date
June 8, 2021
Manufacturer
CROSSROADS EXTREMITY SYSTEMS
Product Code
HRS
PMA / PMN Number
K193452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SURGEON REPORTS THAT THE ANTI-DRIFT BOLT BACKED OUT AND WAS ASYMPTOMATIC ON (B)(6) 2021. THE SCREW CONTINUED TO BACK-OUT AND BECAME SYMPTOMATIC ON (B)(6) 2021. REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 WHERE THE ANTI-DRIFT BOLT WAS REMOVED AND REPLACED. PATIENT RECOVERED AND RESOLVED SYMPTOMS ASSOCIATED WITH LAPIDUS PROCEDURE. PATIENT REPORTS ANKLE POPPING AND 5TH TOE SPASMS LIKELY UNRELATED. NON-CONFORMANCE WAS NOT IDENTIFIED BASED ON THE LOTS THAT WOULD CONTRIBUTE TO THESE EVENTS. SCREW BACK-OUT MAY HAVE BEEN DUE TO INADEQUATE FIXATION INTRA-OPERATIVELY. THE FOLLOWING SCREWS AND PLATES WERE IMPLANTED WITH THE ADB AND NOT REVISED OR REMOVED: REF: 1500-3518 LN: 400074 QTY: 1 PRODUCT NAME: MOTOBAND CP NON-LOCKING SCREWS 3.5MM X 18MM. REF: 7118-1818KT LN: 300288 QTY: 1 PRODUCT NAME: IMPLANT KIT, DYNAFORCE HIMAX, 18MM X 18MM X 18MM. REF: 1500-3514 LN: 400147 QTY: 2 PRODUCT NAME: MOTOBAND CP - NONLOCKING SCREWS 3.5MM X 14MM. REF: 7100-LP18-R LN: 501209 QTY: 1 PRODUCT NAME: DYNABUNION PLATE RIGHT 18MM.

Description of Event or Problem · 1

THE SURGEON REPORTS THAT THE ANTI-DRIFT BOLT BACKED OUT AND WAS ASYMPTOMATIC ON (B)(6) 2021. THE SCREW CONTINUED TO BACK-OUT AND BECAME SYMPTOMATIC ON (B)(6) 2021. REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 WHERE THE ANTI-DRIFT BOLT WAS REMOVED AND REPLACED. PATIENT RECOVERED AND RESOLVED SYMPTOMS ASSOCIATED WITH LAPIDUS PROCEDURE. PATIENT REPORTS ANKLE POPPING AND 5TH TOE SPASMS LIKELY UNRELATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
855040 MOTOBAND DYNAFORCE DYNABUNION PLATE, FIXATION, BONE HRS CROSSROADS EXTREMITY SYSTEMS UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 32 YR Required Intervention