FDA Adverse Event Malfunction Summary report: N

ARCTIC GEL PADS

MDR report key: 11958691 · Received June 8, 2021

Report

Report Number
1018233-2021-03313
Event Type
Malfunction
Date Received
June 8, 2021
Date of Event
April 4, 2021
Report Date
August 29, 2021
Manufacturer
MEDIVANCE, INC. ¿ 1725056
Product Code
DWJ
PMA / PMN Number
K142702
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS INCONCLUSIVE. IT WAS UNKNOWN WHETHER THIS DEVICE, USED FOR TREATMENT, MET SPECIFICATIONS SINCE A DEVICE WAS NOT RECEIVED. NO SAMPLE WAS RETURNED FOR EVALUATION. A POTENTIAL FAILURE MODE COULD BE ¿HYDROGEL PEELS AWAY FROM FILM¿. A POTENTIAL ROOT CAUSE FOR THIS FAILURE COULD BE "IMPROPER DESIGN CONSIDERATION OR MATERIAL SELECTION". THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "DIRECTIONS FOR USE 1. ARCTICGEL¿ PADS ARE ONLY FOR USE WITH AN ARCTIC SUN® TEMPERATURE MANAGEMENT SYSTEM CONTROL MODULE. SEE OPERATORS MANUAL FOR DETAILED INSTRUCTIONS ON SYSTEM USE. 2. SELECT THE PROPER NUMBER, SIZE AND STYLE PAD FOR THE PATIENT SIZE AND CLINICAL INDICATION. HOWEVER, THE RATE OF TEMPERATURE CHANGE AND POTENTIALLY THE FINAL ACHIEVABLE TEMPERATURE IS AFFECTED BY PAD SURFACE AREA, PATIENT SIZE, PAD PLACEMENT AND WATER TEMPERATURE RANGE. BEST SYSTEM PERFORMANCE WILL BE ACHIEVED BY USING THE ENTIRE PAD SET (4). IF THE ENTIRE SET OF PADS IS NOT USED, THE MINIMUM FLOW RATE MAY NOT BE ACHIEVED. 3. FOR PATIENT COMFORT, THE PADS MAY BE PREWARMED USING WATER TEMPERATURE CONTROL MODE (MANUAL) PRIOR TO APPLICATION. 4. PLACE THE PADS ON HEALTHY, CLEAN SKIN ONLY. REMOVE ANY CREAMS OR LOTIONS FROM PATIENT¿S SKIN BEFORE PAD APPLICATION. REMOVE THE RELEASE LINER FROM EACH PAD AND APPLY TO THE APPROPRIATE AREA. THE PADS MAY BE OVERLAPPED OR FOLDED ADHESIVE-TO-ADHESIVE TO ACHIEVE PROPER PLACEMENT. THE PADS MAY BE REMOVED AND REAPPLIED IF NECESSARY. THE PAD SURFACE MUST BE CONTACTING THE SKIN FOR OPTIMAL ENERGY TRANSFER EFFICIENCY. PLACE PADS TO ALLOW FOR FULL RESPIRATORY EXCURSION. 5. ATTACH THE PAD¿S LINE CONNECTORS TO THE PATIENT LINE MANIFOLDS. BEGIN CIRCULATING WATER THROUGH THE PADS USING EITHER PATIENT TEMPERATURE CONTROL MODE (AUTOMATIC) OR WATER TEMPERATURE CONTROL MODE (MANUAL). IF THE PADS FAIL TO PRIME OR A SIGNIFICANT CONTINUOUS AIR LEAK IS OBSERVED IN THE PAD RETURN LINE, CHECK CONNECTIONS, THEN IF NEEDED REPLACE THE LEAKING PAD. 6. ONCE THE PADS ARE PRIMED, ASSURE THE FLOW RATE DISPLAYED ON THE CONTROL PANEL IS GREATER THAN 1.7 LITERS PER MINUTE, WHICH IS THE MINIMUM FLOW RATE FOR A FULL PAD KIT (4). 7. WHEN FINISHED, EMPTY WATER FROM PADS. COLD TEMPERATURE INCREASES THE ADHESIVENESS OF THE HYDROGEL. FOR EASE OF REMOVAL, LEAVE PADS ON THE PATIENT FOR APPROXIMATELY 15 MINUTES TO ALLOW THE HYDROGEL TO WARM. SLOWLY REMOVE PADS FROM THE PATIENT AND DISCARD. ANY SERIOUS INCIDENT THAT HAS OCCURRED IN RELATION TO THE DEVICE SHOULD BE REPORTED TO THE MANUFACTURER AND THE COMPETENT AUTHORITY OF THE MEMBER STATE IN WHICH THE USER AND/OR PATIENT IS ESTABLISHED." H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT REQUIRED TARGETED TEMPERATURE MANAGEMENT POST ARREST. ALSO STATED THAT LEG PAD ADHESIVE PEELED OFF OF THE PAD. PER FOLLOW UP ON (B)(6) 2021, PAD WAS REPLACED AND DISPOSED. PATIENT COMPLETED THERAPY WITH NEW PAD.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT REQUIRED TARGETED TEMPERATURE MANAGEMENT POST ARREST. ALSO STATED THAT LEG PAD ADHESIVE PEELED OFF OF THE PAD. PER FOLLOW UP ON (B)(6) 2021, PAD WAS REPLACED AND DISPOSED. PATIENT COMPLETED THERAPY WITH NEW PAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
852684 ARCTIC GEL PADS ARCTIC GEL PADS DWJ MEDIVANCE, INC. ¿ 1725056 NGFN1444

Patients

Seq Age Sex Outcome Treatment
1 Other