FDA Adverse Event Malfunction Summary report: N

ATTUNE PATELLA COMPRESSION RNG

MDR report key: 11957874 · Received June 8, 2021

Report

Report Number
1818910-2021-12054
Event Type
Malfunction
Date Received
June 8, 2021
Date of Event
May 24, 2021
Report Date
May 26, 2021
Manufacturer
DEPUY IRELAND - 9616671
Product Code
HWN
UDI-DI
10603295395454
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # : (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT (B)(4). INVESTIGATION SUMMARY : EXAMINATION OF THE SUBMITTED DEVICE REVEALED THE RUBBER RING WAS DAMAGE, BUT NOT BROKEN INTO TWO PIECES. THE RUBBER OF THE COMPRESSION RING WAS FOUND TO BE SCRATCHED, NICKED AND CRACKED. THE INVESTIGATION DID NOT INDICATED THAT A BROADER INVESTIGATION OR CORRECTIVE ACTION WAS NECESSARY. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT :THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). WHERE THE PRODUCT AND LOT CODE WAS PROVIDED, A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE ORANGE WASHERS FROM THE PATELLA CLAMPS BROKE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE BROKEN INSTRUMENTS FROM CONSIGNMENT ATTUNE TRAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
858701 ATTUNE PATELLA COMPRESSION RNG ATTUNE INSTRUMENTS : PATELLA HANDLE/CLAMPS HWN DEPUY IRELAND - 9616671 2545-01-042 10603295395454

Patients

Seq Age Sex Outcome Treatment
1