ATTUNE PATELLA COMPRESSION RNG
Report
- Report Number
- 1818910-2021-12051
- Event Type
- Malfunction
- Date Received
- June 8, 2021
- Date of Event
- May 24, 2021
- Report Date
- May 25, 2021
- Manufacturer
- DEPUY IRELAND - 9616671
- Product Code
- HWN
- UDI-DI
- 10603295395454
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT COMPLAINT (B)(4). INVESTIGATION SUMMARY: EXAMINATION OF THE SUBMITTED DEVICE REVEALED THE RUBBER RING WAS DAMAGE, BUT NOT BROKEN INTO TWO PIECES. THE RUBBER OF THE COMPRESSION RING WAS FOUND TO BE SCRATCHED, NICKED AND CRACKED. THE INVESTIGATION DID NOT INDICATED THAT A BROADER INVESTIGATION OR CORRECTIVE ACTION WAS NECESSARY. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT :THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). WHERE THE PRODUCT AND LOT CODE WAS PROVIDED, A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: ADDED: D9. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. CORRECTED: H3.
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE ORANGE WASHERS FROM THE PATELLA CLAMPS BROKE.
(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT THERE WERE BROKEN INSTRUMENTS FROM CONSIGNMENT ATTUNE TRAYS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 858702 | ATTUNE PATELLA COMPRESSION RNG | ATTUNE INSTRUMENTS : PATELLA HANDLE/CLAMPS | HWN | DEPUY IRELAND - 9616671 | 2545-01-042 | 10603295395454 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |