FDA Adverse Event Injury Summary report: N

ATTAIN PERFORMA STRAIGHT MRI SURESCAN

MDR report key: 11952359 · Received June 7, 2021

Report

Report Number
2649622-2021-11329
Event Type
Injury
Date Received
June 7, 2021
Date of Event
May 21, 2021
Report Date
May 28, 2024
Manufacturer
MPRI
Product Code
OJX
UDI-DI
00643169002517
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS ADDITIONAL INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. MEDTRONIC WAS MADE AWARE OF THIS EVENT THROUGH A SEARCH OF LITERATURE PUBLICATIONS. REFERENCED ARTICLE: CASE REPORT: LEAD MACRO DISLODGEMENT: AN UNUSUAL CASE OF LATE-ONSET REEL SYNDROME. CJC OPEN. 2024. 6;104-107. DOI: 10.1016/J.CJCO.2023.11.012. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS FURTHER REPORTED VIA A LITERATURE ARTICLE THAT BOTH THE LV AND RA LEADS EXHIBITED AN INCREASE IN THRESHOLDS. RADIOGRAPHS PERFORMED AT THE OFFICE VISIT SHOWED THE MIGRATION OF THE RA AND LV LEADS INTO THE SUPERIOR VENA CAVA - RA JUNCTION AND THE LEADS APPEARED TO BE COILED AROUND THE DEVICE. IN ADDITION, THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAD MIGRATED FROM ITS ORIGINAL LOCATION. THE DEVICE WAS EXPLANTED ALONG WITH THE LEADS DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI).

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: 3093-28 LEAD IMPLANTED: (B)(6) 2015, 3058 IPG IMPLANTED: (B)(6) 2015. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD AND RIGHT ATRIAL (RA) LEAD DISLODGED AND WERE EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
847545 ATTAIN PERFORMA STRAIGHT MRI SURESCAN DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P OJX MPRI 439888 00643169002517

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Hospitalization| R 5076-52 LEAD, DTBA1QQ CRT-D, 6935M55 LEAD