FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 11949456 · Received June 7, 2021

Report

Report Number
1645337-2021-06364
Event Type
Injury
Date Received
June 7, 2021
Date of Event
January 1, 2019
Report Date
May 11, 2021
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001317
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON JUNE 29, 2021, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON JULY 1, 2021, MENTOR BECAME AWARE OF THE FOLLOWING, AND ASSOCIATED FIELDS HAVE BEEN UPDATED: FIELD D4 FOR LOT NUMBER HAS BEEN UPDATED TO "241895". FIELD D4 FOR SERIAL NUMBER HAS BEEN UPDATED TO (B)(6). A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. ON JULY 8, 2021, DEVICE EVALUATION WAS COMPLETED. DEVICE EVALUATION SUMMARY: MENTOR CONDUCTED A VISUAL INSPECTION, LEAK TEST, AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. DURING THE VISUAL EVALUATION, NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SAL SMOOTH RND DIAP 575CC RETURNED DEVICE. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND A TEAR WAS OBSERVED ON THE ANTERIOR ASPECT MEASURING APPROXIMATELY 0.1 CM. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN THE WHOLE AREA OF THE TEAR. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. BASED ON THE INFORMATION CURRENTLY AVAILABLE, MICROSCOPIC EXAMINATION OF THE RETURNED PRODUCT INDICATES THAT THE IMPLANT COULD HAVE BEING DAMAGED DURING OR SUBSEQUENT TO IMPLANTATION. THE PRODUCT INSERT DATA SHEET CAUTIONS TO NOT ALLOW CAUTERY DEVICES OR SHARP INSTRUMENTS, SUCH AS SCALPELS, SUTURE NEEDLES, HYPODERMIC NEEDLES, HEMOSTATS, ADSON FORCEPS OR SCISSORS TO CONTACT THE DEVICE DURING THE IMPLANTATION OR OTHER SURGICAL PROCEDURES. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: LEFT DEFLATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) BLACK OR AFRICAN AMERICAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH A 575CC MENTOR SMOOTH ROUND MODERATE PROFILE AND EXPERIENCED BREAST IMPLANT DEFLATION, AND A POCKET THAT FELT LIKE A BUBBLE ON HER LEFT SIDE POSTOPERATIVELY. THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH CATALOG NUMBER 3503380 ON (B)(6) 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
842653 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3501690 241895 00081317001317

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention