FDA Adverse Event Malfunction Summary report: N

LUMIRADX SARS-COV-2 AG TEST STRIP KIT

MDR report key: 11949227 · Received June 7, 2021

Report

Report Number
3012642695-2021-01147
Event Type
Malfunction
Date Received
June 7, 2021
Date of Event
May 20, 2021
Report Date
June 7, 2021
Manufacturer
LUMIRADX UK LTD.
Product Code
QKP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT WAS REPORTED TO BE ASYMPTOMATIC WHEN INITIAL LUMIRADX SARS-COV-2 AG TEST WAS PERFORMED ON (B)(6) 2021 WITH STRIP LOT 5000304 AND INSTRUMENT SERIAL NUMBER (B)(4). CONFIRMATORY TESTING WAS PERFORMED VIA POLYMERASE CHAIN REACTION (PCR). THE CONFIRMATORY TEST RESULT WAS NEGATIVE. THE CUSTOMER REPORTED THE FOLLOWING PROCESS FOR TESTING ON THE LUMIRADX PLATFORM: "SINGLE TEST COLLECTED AND PREPARED AT A TIME." THE CUSTOMER REPORTED THEIR CLEANING PRACTICES TO BE "ALCOHOL 70%", AND FURTHER REPORTED THAT GLOVES ARE CHANGED AFTER EACH SAMPLE. LOT 5000304 MET ALL DEFINED QC CRITERIA AT THE TIME IT WAS RELEASED AND TESTING USING NEGATIVE NASAL SWABS FROM IN-HOUSE DONORS MEETS EXPECTED PERFORMANCE FOR USE IN THE FIELD. REVIEW OF PRODUCT RISK ASSESSMENT - SARS-COV-2 AG ASSAY REVISION 7, RESULTED IN SEVERITY OF MINOR, AS FOLLOWS: ASYMPTOMATIC PATIENTS COULD EXPERIENCE SECONDARY HARM OF UNNECESSARY SELF-ISOLATION AND POSSIBLE STRESS/ANXIETY. NO PATIENT HARM, INJURY OR ADVERSE HEALTH CONSEQUENCES WERE COMMUNICATED BY THE CUSTOMER TO LUMIRADX FOR THE REPORTED DISCORDANT RESULT. TRENDING DATA FOR DISCORDANT RESULTS WAS REVIEWED FOR THIS LOT AND THE OCCURRENCE RATE PER QUANTITY OF STRIPS IN THE FIELD WAS CALCULATED AS (B)(4). LUMIRADX SARS-COV-2 AG TEST PRODUCT INSERT CLAIMS A SPECIFICITY OF 96.6% WITH A REFERENCE RT-PCR ASSAY AND IT IS ACCEPTED THAT UP TO 3.4% OF TEST STRIPS MAY GENERATE A DISCORDANT FALSE POSITIVE RESULT. ROOT CAUSE DETERMINATION: NO DEFINITIVE ROOT CAUSE HAS BEEN DETERMINED FOR THE REPORTED DISCORDANT RESULT BASED ON THE INFORMATION CURRENTLY AVAILABLE. INVESTIGATION CONCLUSION AND STATUS OF INVESTIGATION: NO FURTHER INVESTIGATION IS CONSIDERED NECESSARY AT THIS TIME. THIS STRIP LOT CONTINUES TO DEMONSTRATE FIELD PERFORMANCE WITHIN SPECIFICATION OF PRODUCT CLAIMS RELATIVE TO THE QUANTITY OF STRIPS FROM THIS LOT IN THE FIELD. REPORTS OF DISCORDANT RESULTS FOR THIS LOT WILL CONTINUE TO BE TRENDED AND REVIEWED, WITH ACTION TAKEN AS APPROPRIATE IN RESPONSE TO ANY ADVERSE TRENDS OR EVENTS INCLUDING A FOLLOW-UP REPORT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A SUSPECTED FALSE (INCORRECT) POSITIVE RESULT FROM A RAPID RESULT COVID TEST SYSTEM ON AN INDIVIDUAL ASYMPTOMATIC PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
844394 LUMIRADX SARS-COV-2 AG TEST STRIP KIT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM QKP LUMIRADX UK LTD. 5000304

Patients

Seq Age Sex Outcome Treatment
1