FDA Adverse Event Injury Summary report: N

RESPIRONICS

MDR report key: 11946304 · Received June 7, 2021

Report

Report Number
2031642-2021-03975
Event Type
Injury
Date Received
June 7, 2021
Date of Event
May 10, 2021
Manufacturer
RESPIRONICS CALIFORNIA, LLC
Product Code
MNT
UDI-DI
00884838020054
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

DUE DILIGENCE AND MULTIPLE GOOD FAITH EFFORTS HAVE BEEN MADE TO OBTAIN ADDITIONAL RELEVANT INFORMATION WITH NO RESPONSE FROM THE CUSTOMER. CLINICAL DETAILS SURROUNDING THE EVENT OF INTUBATION, PATIENT OUTCOME, DEVICE RESOLUTION AND DISPOSITION, CONCLUSION AND ROOT CAUSE REMAIN UNKNOWN DUE TO LACK OF INFORMATION. THE COMPLAINT WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED AT A LATER DATE, THE COMPLAINT WILL BE REOPENED, AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

B4:27AUG2021. NEW INFORMATION WAS RECEIVED. THE DEVICE WAS IN USE ON A PATIENT; THE DEVICE STOPPED AND THE PATIENT NEEDED TO BE INTUBATED. THERE WAS A DELAY IN PROVIDING THERAPY. THE DIAGNOSTIC REPORT NOTES PRESSURE REGULATION HIGH ERROR OCCURRED ON (B)(6) 2021. ON (B)(6) 2021, A PHILIPS AUTHORIZED SERVICE PERSONNEL (ASP) WAS DISPATCHED TO THE CUSTOMER SITE TO IMPLEMENT FIELD CHANGE ORDER FOR THE POWER MANAGEMENT BOARD. THE PM PCBA WAS REPLACED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Additional Manufacturer Narrative · 1

DATE OF REPORT: 07JUN2021.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE UNIT HAS ERROR MESSAGE- INOP CALL FOR SERVICE. THE DEVICE STOPPED ON A PATIENT. THE CUSTOMER CONTACTED PRODUCT SUPPORT AND REQUESTED FOR SERVICE REPAIR. THE CUSTOMER REPORTED THAT THE UNIT WAS IN USE ON PATIENT, BUT THERE WAS NO PATIENT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
847224 RESPIRONICS VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE MNT RESPIRONICS CALIFORNIA, LLC V60 00884838020054

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention