FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER SST II ADVANCE PLUS BLOOD COLLECTION TUBES

MDR report key: 11942339 · Received June 4, 2021

Report

Report Number
9617032-2021-00594
Event Type
Malfunction
Date Received
June 4, 2021
Date of Event
May 11, 2021
Report Date
May 18, 2021
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 2 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, 4 RETENTION SAMPLES FROM BD INVENTORY WERE DRAWN WITH HORSE BLOOD, MIXED, AND STOOD FOR 30 MINUTES. THEY WERE THEN CENTRIFUGED AT 1211RCF FOR 10 MINUTES, USING AN MSE MISTRAL 1000 CENTRIFUGE. ALL SAMPLES SEPARATED AS EXPECTED WITH COMPLETE IMPERVIOUS GEL BARRIERS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED BASED ON THE REVIEW OF THE PHOTOS. A ROOT CAUSE COULD NOT BE DETERMINED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED POOR BARRIER SEPARATION OF THE SAMPLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "SST TUBES CENTRIFUGED AT 2200G FOR 10 MINUTES AND GEL REMAINED AT THE BASE OF THE TUBE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
833952 BD VACUTAINER SST II ADVANCE PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 0240259

Patients

Seq Age Sex Outcome Treatment
1