FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11941042 · Received June 4, 2021

Report

Report Number
1221359-2021-01693
Event Type
Malfunction
Date Received
June 4, 2021
Date of Event
March 17, 2021
Report Date
November 22, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
Product Code
QJR
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1017966 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1017966 , TEST BASE PART NUMBER 190-430 / LOT: 1017966 . THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1017966 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED.

Additional Manufacturer Narrative · 1

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED. REFERENCE MFR REPORTS: 1221359-2021-01682, 1221359-2021-01692, 1221359-2021-01694.

Description of Event or Problem · 1

THE CUSTOMER REPORTED 4 FALSE POSITIVE RESULTS ON DIRECT TESTED NASOPHARYNGEAL COPAN SWABS WITH THE ID NOW COVID-19 ASSAY ON (B)(6) 2021. THIS MFR. ADDRESSES PATIENT 3 OF 4. CONFIRMATION TESTING WAS PERFORMED ON THE SAME SAMPLE AND GENERATED NEGATIVE RESULTS. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO TEST RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
833869 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC 1017966

Patients

Seq Age Sex Outcome Treatment
1 4 YR Female