FDA Adverse Event Injury Summary report: N

MONARCH ENDOSCOPY PLATFORM (MONARCH PLATFORM)

MDR report key: 11939899 · Received June 4, 2021

Report

Report Number
3014447948-2021-00017
Event Type
Injury
Date Received
June 4, 2021
Date of Event
May 7, 2021
Report Date
May 7, 2021
Manufacturer
AURIS HEALTH, INC.
Product Code
EOQ
PMA / PMN Number
K173760
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CORRECTED MANUFACTURER ADDRESS STREET: 150 SHORELINE DRIVE MANUFACTURER ADDRESS STREET: 150 SHORELINE DRIVE.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THERE WAS NO ALLEGATION OF DEVICE FAILURE. RISK OF PNEUMOTHORAX IS DOCUMENTED IN 105-000299-01, REV D, MONARCH BRONCH RISK MANAGEMENT REPORT AND IS DEEMED TO BE ACCEPTABLE. BASED ON THE INFORMATION AVAILABLE FOR THIS COMPLAINT, THE ROOT CAUSE OF THE REPORTED EVENT IS RELATED TO A KNOWN INHERENT RISK OF THE DEVICE AND PROCEDURE AS DOCUMENTED IN THE LABELING. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. THE REPORTED EVENT WILL CONTINUE TO BE MONITORED THROUGH REGULAR COMPLAINT TREND REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A MONARCH-ASSISTED BRONCHOSCOPY PROCEDURE THE PATIENT EXPERIENCED A PNEUMOTHORAX. THE LOCATION WAS RIGHT UPPER LOBE. THE LOCATION OF THE PNEUMOTHORAX WAS RIGHT SIDE. DURING THE PROCEDURE THE PATIENT WENT INTO A COMPLETE HEART BLOCK AND BECAME HYPOTENSIVE. EPINEPHRINE WAS PUSHED INTO THE PATIENT¿S INTRAVENOUS TO TACKLE HYPOTENSION AND THE SCOPE WAS REMOVED IMMEDIATELY. THE CUSTOMER TOOK THE SCOPE OFF THE ROBOTIC ARMS AT THIS TIME AND WENT ON TO REMOVE THE SCOPE OUT OF PATIENT MANUALLY. THE PATIENT SUBSEQUENTLY BECAME STABLE, E.G., SHOWED INFERIOR WALL ACUTE MYOCARDIAL INFARCTION (AMI). THE PATIENT WENT TO THE CARDIAC CATHETERIZATION LAB IMMEDIATELY AND HAD AN EMERGENT ANGIOGRAM. THE PATIENT HAS RECOVERED AND HAS BEEN DISCHARGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
837795 MONARCH ENDOSCOPY PLATFORM (MONARCH PLATFORM) BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES EOQ AURIS HEALTH, INC. MON-000005

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R