FDA Adverse Event
Summary report: N
IRIX 70
MDR report key: 1193545
·
Received October 8, 2008
Report
- Report Number
- 8020825-2008-00003
- Date Received
- October 8, 2008
- Date of Event
- July 23, 2008
- Report Date
- August 10, 2008
- Manufacturer
- TROPHY RADIOLOGIE
- Product Code
- EAP
- Removal / Correction Number
- Z-1099/1101-5
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED CONDITION IS ATTRIBUTED TO A FAILURE OF THE SUPPORT ARM. THIS CONDITION WAS IDENTIFIED BY THE MFR IN 1995 AND CORRECTIVE ACTIONS WERE INSTITUTED. THE DEVICE CITED IN THIS REPORT IS INVOLVED IN RECALL Z-1099/1101-5.
Description of Event or Problem · 1
IRIX SCISSOR ARM BROKE AT KNUCKLE. ARM FELL AND STRUCK PT ON OR ABOUT HEAD. MEDICAL ATTENTION WAS OBTAINED. PT NAME NOT DISCLOSED.