LIFESHIELD SYMBIQ SET WITH 2 CLAVES
Report
- Report Number
- 9615050-2008-00292
- Event Type
- Malfunction
- Date Received
- October 7, 2008
- Date of Event
- September 17, 2008
- Report Date
- September 17, 2008
- Manufacturer
- HOSPIRA DE COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K041550
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
DUE TO HAZARDOUS CONTAMINATION, THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION.
THE CUSTOMER CONTACT REPORTED BACKFLOW OF FLUID FROM THE SECONDARY TUBING SET INTO THE PRIMARY TUBING SET. THE PRIMARY TUBING SET WAS BEING USED TO DELIVER NORMAL SALINE VIA A SYMBIQ PUMP. THE SECONDARY TUBING SET WAS CONNECTED TO THE PRIMARY TUBING SET FOR PIGGYBACK DELIVERY OF AN UNSPECIFIED CONCENTRATION OF DOXORUBICIN IN A 500ML SOLUTION CONTAINER. THE PRIMARY SOLUTION CONTAINER WAS LOWERED WITH "2 HANGERS" BELOW THE SECONDARY SOLUTION CONTAINER; HOWEVER, THE NURSE NOTED THE SECONDARY SOLUTION WAS FLOWING INTO THE PRIMARY TUBING SET. IT WAS REPORTED THE PRIMARY TUBING SET WAS CLAMPED BELOW THE BACK CHECK VALVE AND THE THERAPY WAS RESUMED. THERE WERE NO REPORTED ADVERSE PT EFFECTS. NO MEDICAL INTERVENTIONS WERE REQUIRED. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFESHIELD SYMBIQ SET WITH 2 CLAVES | 80-FRN | FRN | HOSPIRA DE COSTA RICA LTD. | NA | 631505H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LIST # 11953| LOT # UNK| LIFESHIELD SECONDARY TUBING SET |