FDA Adverse Event Malfunction Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 11934365 · Received June 3, 2021

Report

Report Number
1221359-2021-01643
Event Type
Malfunction
Date Received
June 3, 2021
Report Date
August 13, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORT : 1221359-2021-01640, 1221359-2021-01641, 1221359-2021-01642.

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 1014263 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: 1014263, TEST BASE PART NUMBER 190-430 / LOT: 1014263 THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1014263 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED SEVEN (7) FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY, WHICH OCCURRED ON VARIOUS DATES AND WITH DIFFERENT LOT NUMBERS. THIS REPORT ADDRESSES LOT FOUR (4) OF FOUR (4). THE CUSTOMER REPORTED ONE (1) FALSE NEGATIVE RESULT WITH THE ID NOW COVID-19 ASSAY . THE NASOPHARYGAL SAMPLE WAS COLLECTED ON (B)(6) 2021 . REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING (CORELAB) GENERATED POSITIVE RESULTS. THE NASOPHARYNGEAL SAMPLE IN VIRAL TRANSPORT MEDIA WAS COLLECTED ON (B)(6) 2021. PER THE CUSTOMER, NO ADDITIONAL INFORMATION WILL BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
829825 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1014263

Patients

Seq Age Sex Outcome Treatment
1