ID NOW COVID 19 ASSAY
Report
- Report Number
- 1221359-2021-01642
- Event Type
- Malfunction
- Date Received
- June 3, 2021
- Report Date
- August 13, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H10: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1020568 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: 1020568, TEST BASE PART NUMBER 190-430 / LOT: 1020568 THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1020568 SHOWED THAT THE COMPLAINT RATE IS 0.015%. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED.
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORT : 1221359-2021-01640, 1221359-2021-01641, 1221359-2021-01643.
THE CUSTOMER REPORTED SEVEN (7) FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY, WHICH OCCURRED ON VARIOUS DATES AND WITH DIFFERENT LOT NUMBERS. THIS REPORT ADDRESSES LOT THREE (3) OF FOUR (4). THE CUSTOMER REPORTED TWO (2) FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY. NASAL SWABS WERE COLLECTED ON (B)(6) 2021. CONFIRMATION TESTING (B)(4) GENERATED POSITIVE RESULTS. THE NASOPHARYNGEAL SAMPLES IN VIRAL TRANSPORT MEDIA WERE COLLECTED ON (B)(6) 2021. PER THE CUSTOMER, NO ADDITIONAL INFORMATION WILL BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 827674 | ID NOW COVID 19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 1020568 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |