FDA Adverse Event Malfunction Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 11934317 · Received June 3, 2021

Report

Report Number
1221359-2021-01642
Event Type
Malfunction
Date Received
June 3, 2021
Report Date
August 13, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H10: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1020568 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: 1020568, TEST BASE PART NUMBER 190-430 / LOT: 1020568 THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1020568 SHOWED THAT THE COMPLAINT RATE IS 0.015%. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORT : 1221359-2021-01640, 1221359-2021-01641, 1221359-2021-01643.

Description of Event or Problem · 1

THE CUSTOMER REPORTED SEVEN (7) FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY, WHICH OCCURRED ON VARIOUS DATES AND WITH DIFFERENT LOT NUMBERS. THIS REPORT ADDRESSES LOT THREE (3) OF FOUR (4). THE CUSTOMER REPORTED TWO (2) FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY. NASAL SWABS WERE COLLECTED ON (B)(6) 2021. CONFIRMATION TESTING (B)(4) GENERATED POSITIVE RESULTS. THE NASOPHARYNGEAL SAMPLES IN VIRAL TRANSPORT MEDIA WERE COLLECTED ON (B)(6) 2021. PER THE CUSTOMER, NO ADDITIONAL INFORMATION WILL BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
827674 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1020568

Patients

Seq Age Sex Outcome Treatment
1