FDA Adverse Event Malfunction Summary report: N

TRANSEND GUIDEWIRE

MDR report key: 1193422 · Received October 8, 2008

Report

Report Number
2939204-2008-00454
Event Type
Malfunction
Date Received
October 8, 2008
Date of Event
June 16, 2008
Report Date
September 16, 2008
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
DQX
PMA / PMN Number
K022357
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE GUIDEWIRE (SUBJECT DEVICE) WAS RETURNED TO BOSTON SCIENTIFIC. UPON VISUAL EXAMINATION PERFORMED ON THE GUIDEWIRE, IT WAS OBSERVED THAT THE SPRING TIP WAS FORMED INTO A LOOP AND ELONGATED APPROXIMATELY 2 MM FROM THE PROXIMAL SOLDER POINT. THE WIRE WAS FOUND TO BE KINKED ALONG ITS ENTIRE LENGTH. FURTHER OBSERVATION FOUND THAT THE PTFE COATING WAS SCRAPPED APPROXIMATELY 134 CM THROUGH 140 CM FROM THE DISTAL TIP OF THE SUBJECT DEVICE. THERE WERE NO FURTHER ANOMALIES NOTED TO THE SUBJECT DEVICE. NO FURTHER INFO HAS BEEN RECEIVED TO INDICATE THAT THE DEVICE WAS NOT USED AS INSTRUCTED IN THE DIRECTIONS FOR USE. THE DEVICE HISTORY RECORD REVIEW DID NOT INDICATE ANY ISSUES OR DISCREPANCIES DURING THE MANUFACTURING PROCESS OF THE RETURNED GUIDEWIRE THAT COULD HAVE CONTRIBUTED TO EITHER THE REPORTED ISSUES OR OBSERVED ANOMALIES. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION IN REGARDS TO THE SUBJECT DEVICE; HOWEVER, NO RESPONSE WAS RECEIVED BACK FROM THE FACILITY. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED BASED ON THE LIMITED INFORMATION AVAILABLE. IT IS PROBABLE THAT THE DAMAGE TO THE COATING OCCURRED FOLLOWING THE WIRE KINKING; THEREFORE, A ROOT CAUSE OF OPERATIONAL CONTEXT WAS ASSIGNED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC THAT THE SUBJECT GUIDEWIRE KINKED DURING THE PROCEDURE. ANALYSIS OF THE RETURNED DEVICE NOTED THE POLYTETRAFLUOROETHYLENE (PTFE) COATING WAS PEELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSEND GUIDEWIRE (DQX) WIRE, GUIDE, CATHETER DQX BOSTON SCIENTIFIC CORPORATION 46-815 11434276

Patients

Seq Age Sex Outcome Treatment
1