FDA Adverse Event Malfunction Summary report: N

SYNCHRO 2 GUIDEWIRE

MDR report key: 1193418 · Received October 8, 2008

Report

Report Number
2939204-2008-00452
Event Type
Malfunction
Date Received
October 8, 2008
Date of Event
August 20, 2008
Report Date
October 3, 2008
Manufacturer
NEUROVASCULAR DIVISION, BOSTON SCIENTIFIC CORPORATION
Product Code
DQX
PMA / PMN Number
K053268
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION RESULTS: OTHER FOR HANDLING DAMAGE. THE DEVICE HISTORY RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT THE POLYTETRAFLUOROETHYLENE (PTFE) COATING WAS MISSING 18.4CM FROM THE PROXIMAL END, THERE WAS A SLIGHT BEND 34.5CM FROM THE DISTAL END AND THE DISTAL TIP HAD SEPARATED APPROXIMATELY 3.6CM FROM THE DISTAL END. THE DISTAL TIP WAS NOT RETURNED. NO OTHER ANOMALIES WERE FOUND ON THE DEVICE. THE DEVICE WAS LOADED INTO A DEMONSTRATION MICROCATHETER PROXIMAL END AND ADVANCED TO THE DISTAL END WITHOUT ANY DIFFICULTY. THE DAMAGE TO THE COATING AND THE KINK ARE INDICATIVE OF RESISTANCE/JAMMING BETWEEN THE DEVICE AND MICROCATHETER. THE DEFINITIVE CAUSE FOR THE RESISTANCE ENCOUNTERED AND THE SEPARATION OF THE DISTAL TIP CANNOT BE DETERMINED. FROM ADDITIONAL INFORMATION PROVIDED, IT WAS DETERMINED THAT THERE WAS NO DAMAGE OBSERVED TO THE DEVICE OR MICROCATHETER DURING PREPARATION AND THE DEVICES WERE PREPARED IN ACCORDANCE TO THEIR RESPECTIVE DIRECTIONS FOR USE. THE OBSERVED DAMAGES MOST LIKELY OCCURRED DURING PREPARATION OR, AS A RESULT OF THE REPORTED RESISTANCE ENCOUNTERED DURING THE PREPARATION TO USE THE DEVICES IN THE PROCEDURE; SO, A ROOT CAUSE OF HANDLING DAMAGE WAS ASSIGNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE PREPARATION, AS THE GUIDEWIRE WAS BEING LOADED INTO THE MICROCATHETER, IT BECAME "EFFECTIVELY STUCK" IN THE MICROCATHETER AND THE PHYSICIAN WAS UNABLE TO ADVANCE OR RETRACT THE GUIDEWIRE. THE DEVICES WERE NOT USED IN THE PROCEDURE. EVALUATION OF THE RETURNED GUIDEWIRE FOUND THAT PART OF THE DISTAL TIP HAD SEPARATED FROM THE DEVICE AND THE COATING WAS PEELING AT THE PROXIMAL END.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRO 2 GUIDEWIRE (DQX) GUIDEWIRE DQX NEUROVASCULAR DIVISION, BOSTON SCIENTIFIC CORPORATION 2601 B14178

Patients

Seq Age Sex Outcome Treatment
1 BOSTON SCIENTIFIC EXCELSIOR SL10 MICROCATHETER