COBAS INTEGRA 800
MDR report key: 1193376
·
Received October 7, 2008
Report
- Report Number
- 1823260-2008-07401
- Event Type
- Malfunction
- Date Received
- October 7, 2008
- Date of Event
- September 15, 2008
- Report Date
- October 7, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K951595
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
USER EXPERIENCED DISCREPANT URIC ACID RESULTS FOR ONE PATIENT SAMPLE. INITIAL RESULT GAVE 0 MG/DL; REPEATED TWICE GAVE 4.9 AND 4.8 MG/DL. THE ORIGINAL RESULT WAS NOT REPORTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE A BROKEN LIQUID LEVEL DETECTION CABLE, AND REPLACED CABLE. PERFORMANCE TESTS WERE PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS INTEGRA 800 | CLINICAL CHEMISTRY ANALYZER - JJE | JJE | ROCHE DIAGNOSTICS | I800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR |