FDA Adverse Event Malfunction Summary report: N

COBAS INTEGRA 800

MDR report key: 1193376 · Received October 7, 2008

Report

Report Number
1823260-2008-07401
Event Type
Malfunction
Date Received
October 7, 2008
Date of Event
September 15, 2008
Report Date
October 7, 2008
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K951595
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

USER EXPERIENCED DISCREPANT URIC ACID RESULTS FOR ONE PATIENT SAMPLE. INITIAL RESULT GAVE 0 MG/DL; REPEATED TWICE GAVE 4.9 AND 4.8 MG/DL. THE ORIGINAL RESULT WAS NOT REPORTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE A BROKEN LIQUID LEVEL DETECTION CABLE, AND REPLACED CABLE. PERFORMANCE TESTS WERE PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS INTEGRA 800 CLINICAL CHEMISTRY ANALYZER - JJE JJE ROCHE DIAGNOSTICS I800

Patients

Seq Age Sex Outcome Treatment
1 86 YR