TRIAGE CARDIAC HS PANEL
Report
- Report Number
- 3013982035-2021-00011
- Event Type
- Malfunction
- Date Received
- June 3, 2021
- Date of Event
- May 3, 2021
- Report Date
- May 4, 2021
- Manufacturer
- QUIDEL CARDIOVASCULAR INC.
- Product Code
- MMI
- PMA / PMN Number
- K030286
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION CONCLUSION: THE CUSTOMERS COMPLAINT WAS NOT REPLICATED DURING IN-HOUSE TESTING OF RETAIN DEVICE LOT T12165N. RETAINS OF THE COMPLAINT LOT PERFORMED PROPERLY WHEN TESTED WITH "NORMAL" (APPARENTLY HEALTHY) DONORS; ALL TNI RESULTS <0.05NG/ML. MANUFACTURING BATCH RECORDS FOR LOT T12165N WERE REVIEWED AND FOUND THAT THE LOT MET FINAL RELEASE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTION ACTION IS REQUIRED.
INVESTIGATION PENDING.
CUSTOMER REPORTED HIGH TROPONINI (TNI) ON TRIAGE WITH ONE PATIENT. ON (B)(6) 2021, (B)(6)-YEAR-OLD MALE PATIENT PRESENTED TO EMERGENCY CENTER WITH HYPERTENSION. INITIAL TRIAGE TEST (DRAWN (B)(6) 2021 7:50AM), TNI = 0.13. RE-DRAW (NO TIME PROVIDED BUT DRAW ON SAME DAY, TEST WAS ORDERED AT 8:56AM, RESULTED AT 10:00AM), RESULTED IN AN ERROR CODE, E1. SAME RE-DRAW SAMPLE REPEAT TESTING, TNI = 0.05. REFERENCE RANGE: 0.0-0.05, CRITICAL (TEST TO BE REPEATED): >/= 0.06. CONFIRMATORY TESTING (POSSIBLY FROM 2ND DRAW USED) TNI = 0.012 (METHOD/INSTRUMENT NOT PROVIDED, REFERENCE RANGE: 0.012-0.033). EKG WAS NORMAL. PATIENT NOT DIAGNOSED WITH MI AND PATIENT NOT ADMITTED. NO PROCEDURAL ISSUES IDENTIFIED. MYO LAB RESULTS: INITIAL=259, RE-DRAW, REPEAT=220.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 831320 | TRIAGE CARDIAC HS PANEL | TRIAGE CARDIAC HS PANEL | MMI | QUIDEL CARDIOVASCULAR INC. | 97000HS | T12165N |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | TRIAGE METERPRO, PN: 55070, SN: (B)(4) |