FDA Adverse Event
Malfunction
Summary report: N
MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP
MDR report key: 1193170
·
Received October 7, 2008
Report
- Report Number
- 1423500-2008-00859
- Event Type
- Malfunction
- Date Received
- October 7, 2008
- Date of Event
- August 7, 2008
- Report Date
- August 12, 2008
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KDJ
- PMA / PMN Number
- K842885A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: RESULTS INDICATE CONFIRMATION OF THE REPORTED EVENT OF A TRANSFER SET'S WHITE AND BLUE PART NOT CLICKING, AND IT'S PLASTIC PEELING OFF. ONE BROKEN OCCLUDER FOOT WAS FOUND DURING EVALUATION. A LIKELY ROOT CAUSE IS THE USE OF ENVIRONMENTAL STRESS CRACKING AGENTS (I.E., CHEMICAL SOLUTIONS IN CONTACT WITH THE PART UNDER AN APPLIED STRESS). RENAL QUALITY ENGINEERING, PLANT MANUFACTURING, AND QUALITY PERSONNEL WILL CONTINUE TO MONITOR THIS PRODUCT FOR TRENDS, AND TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.
Description of Event or Problem · 1
A DISTRIBUTOR REPORTED TO BAXTER THAT A CUSTOMER COMPLAINED THAT A TRANSFER SET'S WHITE AND BLUE PART DOES NOT CLICK ANYMORE, AND THE PLASTIC IS PEELING OFF. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP | 78KDJ | KDJ | BAXTER HEALTHCARE | H07K14044 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |