FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1193149 · Received October 7, 2008

Report

Report Number
2531779-2008-00679
Event Type
Malfunction
Date Received
October 7, 2008
Date of Event
September 8, 2008
Report Date
September 8, 2008
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Removal / Correction Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED AN INTERMITTENT LOSS OF CONTACT BETWEEN THE BATTERY CAP AND THE PUMP CASE.

Description of Event or Problem · 1

PATIENT REPORTS PUMP WAS RESETTING ITSELF WITHOUT USER INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 44 YR