LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1193139
·
Received October 7, 2008
Report
- Report Number
- 3015876-2008-01459
- Event Type
- Malfunction
- Date Received
- October 7, 2008
- Date of Event
- September 5, 2008
- Report Date
- September 7, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT THE ROOT CAUSE OF THE REPORTED FAILURE WAS AN ELECTRICAL SHORT THROUGH FET TRANSISTORS, DESIGNATORS Q1 AND Q2. IT WAS ALSO OBSERVED THAT A FUSE, DESIGNATOR F1, WAS OPEN.
Description of Event or Problem · 1
IT WAS REPORTED THE DEVICE WOULD NOT OPERATE ON AC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.