FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1193081 · Received October 13, 2008

Report

Report Number
6000001-2007-04068
Event Type
Malfunction
Date Received
October 13, 2008
Date of Event
January 1, 2007
Report Date
January 8, 2007
Manufacturer
BAXTER HEALTHCARE PTE LTD
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA FEB 05, 2007. EVALUATION SUMMARY:A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF FAIL CODE 812:02 WAS CONFIRMED IN THE EVENT HISTORY AND ON POWER UP. POSSIBLY, AN UNPLUGGED PUMP HEAD MODULE HARNESS TO MOTORS OR A FAULTY PUMP HEAD MODULE CAUSED THIS FAIL CODE. THE PUMP HEAD MODULE WAS REPLACED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED FAILURE CODE 812:02. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE FAILURE OCCURRED DURING A PATIENT INFUSION. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1